Ole Storstein is first author of a 1959 experimental study separating airway resistance from lung compliance. The investigators recorded mouth-to-esophageal pressure differences, breathing volume and airflow to analyze the mechanical behavior of respiration.[ref]
The apparatus used an intra-esophageal balloon, a differential pressure transducer and a pneumotachograph with volume integration. The paper selected corresponding points of no mouth airflow for its compliance calculation and used paired observations during inflation and deflation to estimate resistance.[ref] Simultaneous signals matter because a pressure observed during flow includes a different mechanical contribution from pressure observed after that flow term vanishes.
For an original one-compartment model with a chosen zero-volume baseline, write
where is lung compliance and is airway resistance. This neglects inertial and regional effects. At stipulated , L, and L/s, the elastic contribution is 4 and the resistive contribution is 1 cm of water. Total modeled pressure is 5 cm of water.
Dividing volume by that total would give , below the chosen compliance. At the same modeled volume and zero flow, pressure becomes 4 cm of water and the ratio recovers 0.10. A time point and the definition of the pressure difference are therefore part of the observable, rather than optional details.
The primary paper explicitly notes the limitation of treating a nonuniform mechanical system as uniform.[ref] Zero net airflow at the mouth need not establish that every region has identical local behavior. Storstein's documented contribution combines apparatus, simultaneous observations and a stated mechanical interpretation, rather than equating one pressure/volume ratio with every aspect of lung function.