Operating-point transfer of an overlap reference asks whether a state-dependent normalization correction measured under reference conditions remains valid at the specimen's optical and electrical settings. Detector bias, optical power, beat frequency, amplifier range and illumination can all define the operating point. Repeated readings at one setting establish repeatability there, rather than transfer over an untested range.
Let the true optical log-ratio be and the instrumental state offset at power be . A reference measurement estimates . Applying it to a specimen at leavesThis follows directly from reference-state transfer in overlap normalization. The difference between operating points is a measurement-model input, not extra independent readout noise.
An original local model writes . References at two powers support linear interpolation at an intermediate :The weights sum to one. A specimen outside the tested range requires extrapolation, which is more sensitive to model error.
Curvature error in overlap-reference transfer is visible in an exactly solvable example. If the true offset is linear plus , its difference from endpoint interpolation is . For powers 1 and 3 mW, suppose the measured endpoint offsets are 0.02 and 0.06. The interpolated offset at 2 mW is 0.04. A quadratic term , with powers expressed in mW, leaves both endpoints unchanged but makes the actual midpoint offset 0.03. Correcting by 0.04 then leaves a log-ratio error of . These are stipulated response values, not observations.
Reference uncertainty also transfers. At a fixed, known interpolation power,Here are standard uncertainties; uncertainty in requires additional terms. A shared calibration can correlate the endpoints, so their covariance should not be dropped merely because they were measured at different powers.[ref]
A useful proposed experiment brackets specimen acquisitions in time, samples several operating powers and repeats at the intended bias and frequency. Check for optical changes as well as electronic response: NIST's primary heterodyne apparatus controls spatial mode and polarization before electrical calibration.[ref]
Frequency response metrology for high-speed optical receivers by Paul D. Hale and others. PDF p. 1, Heterodyne measurements
The interpolation calculation above diagnoses a transfer assumption; it is not an attributed NIST procedure.Curvature error in overlap-reference transfer is the residual instrumental correction left when a curved operating-point response is represented by endpoint linear interpolation. Let be the actual log-ratio offset at optical power , and let interpolate its exact values at . Applying as the correction leavesThis deterministic model discrepancy is separate from random uncertainty in the measured endpoints.
The polynomial interpolation error formula gives, for twice continuously differentiable and between the endpoints,for some in that interval. To derive it, subtract a multiple of chosen to make the error vanish also at ; two applications of Rolle's theorem make the second derivative vanish at an intermediate point. If , thenThe maximum product occurs at the midpoint. Halving the reference spacing therefore quarters this worst-case bound when the same curvature bound remains justified.
For an original example, use power in milliwatts, endpoints , , and . The midpoint error is bounded by in log-ratio. The exact functionattains while matching both endpoint corrections perfectly. The corresponding multiplicative error in the corrected optical ratio is . Repeating endpoint measurements indefinitely would reduce their random uncertainty without revealing this midpoint discrepancy.
To guarantee throughout a reference interval under the stated bound, choose spacing satisfyingThe numerical bound and tolerance require spacing no greater than . Outside the endpoint interval, the product error formula needs a curvature bound over the larger range including the extrapolation point; the interior maximum bound cannot simply be reused.
A proposed check measures additional intermediate power settings and repeats at the intended detector bias and beat frequency. It also monitors mode alignment and polarization, which the primary NIST heterodyne apparatus controls separately.[ref]
Frequency response metrology for high-speed optical receivers by Paul D. Hale and others. PDF p.1, Heterodyne measurements, mode matching
Finitely many extra points constrain a response model but do not prove an arbitrary global curvature bound. State how is justified. The GUM's treatment of uncertainty propagation distinguishes local approximations from significant nonlinear effects.[ref] Articles by others on the same topic