In ideal computed tomography, the unknown is an absorption coefficient . Along a straight ray the transmitted intensity satisfies , soThus calibrated transmission measurements supply the Radon transform of the slice. The inversion above reconstructs the coefficient by filtered backprojection: filter each projection in its transverse variable and then accumulate it over the rays through each image point. This turns a collection of line-integral measurements into a spatial map, which is the key usefulness of the result.
The ideal derivation assumes known ray geometry, sufficiently complete angular sampling and a straight-ray attenuation model. The ramp Fourier multiplier grows at high frequency, so measurement noise is amplified. Practical reconstruction therefore uses discretization and regularization of an inverse problem, for example a window on the ramp filter. Limited angles or missing detector positions cannot be treated as if they were complete noiseless Radon transform data.