Photoconductor rise and fall asymmetry 2026-11-05
Photoconductor rise and fall asymmetry means that a photoconductor's normalized turn-on and turn-off transients have different shapes or threshold-crossing durations under stated illumination and electrical conditions. Define the levels first: a ten-to-ninety-percent rise and a ninety-to-ten-percent fall exclude different portions of a long tail than a zero-to-ninety-percent measurement. Detector temperature is an essential condition because the primary PbS description reports changes in response speed, dark resistance and spectral sensitivity with temperature.[ref]
How does a PbS photoconductive detector work? by Shimadzu Scientific Instruments. temperature-dependence paragraph
The simplest linear carrier model predicts symmetry. For excess carrier density satisfying , a step to gives ; switching off after equilibration gives . Both threshold durations equal . Thus different measured durations are evidence against that complete one-pole measurement model, rather than a result it already predicts.
A photoconductor transient with quadratic recombination supplies an illuminating alternative model. Stipulate constant mobility, fixed detector bias, zero dark excess density andUnder constant , let and . Starting at zero gives . Starting at and setting gives . Direct substitution verifies both solutions.
For an output proportional to , the resulting durations areWith stipulated , the rise takes and the fall , despite having no separate fitted rise and fall constants. This is a derived nonlinear electron-hole recombination model, not measured behavior or a specification of the cited PbS or CdS detector. Its long algebraic tail also shows why a single exponential fit may depend strongly on the fitted interval.
The readout can itself introduce asymmetry. A resistance divider has output as a function of detector electrical conductance . Equal exponential rise and fall in need not give equal normalized threshold durations after this nonlinear conversion. Carrier dynamics, the electrical transfer curve and an amplifier's response must therefore be separated before assigning a material lifetime.
A proposed experiment switches a monitored optical source between two reproducible levels, measures actual detector bias and current, waits for the same initial equilibrium, and repeats in both directions. Compare current-derived conductance with divider output, then vary background, temperature and step size. A slow reference source or readout can imitate a detector tail. Asymmetry alone does not uniquely identify trapping or any other microscopic mechanism.
A photoconductor transient with quadratic recombination is a carrier-density model in which the loss rate grows as the square of the mobile excess density. It explains how nonlinear loss can produce different turn-on and turn-off waveforms without assigning unrelated exponential lifetimes to the two directions. A photoconductor converts illumination-dependent electrical conductivity into a signal under external bias; the manufacturer's primary PbS description establishes that conversion mechanism.[ref]
How does a PbS photoconductive detector work? by Shimadzu Scientific Instruments. band excitation and external resistance readout
The quadratic loss law below is an explicitly stipulated model, not a fitted specification of that detector.Assume a spatially uniform excess density , generation rate , constant coefficient and an output proportional to at fixed mobility and bias:Density has units , has units and has units . A two-carrier recombination rate proportional to the product of equal electron and hole densities motivates a quadratic term in an appropriate regime. Background carriers, other loss processes, contacts and trapping can invalidate that specialization.
For constant , set and . Separating variables, or differentiating the following expression, gives the solution for any finite nonnegative initial density :For it reduces to . For it stays at equilibrium, and for it approaches the same equilibrium from above. In the dark, instead yieldsThe reciprocal-density plot is linear in this model, while a semilogarithmic decay plot is curved. The dark half-decay time is , so it depends on the starting illumination.
The small-signal lifetime under quadratic recombination is different again. Linearize about a nonzero background :A locally exponential small-signal response can therefore coexist with a strongly nonexponential large dark transient. One measured time constant cannot be substituted for all three quantities.
For an original example, stipulate and . Then , dark half-decay time , small-signal lifetime and small-signal half-power frequency about . Four times the background generation doubles equilibrium density and doubles that frequency in this model.
A proposed test measures small modulation about several backgrounds separately from complete turn-off traces, while controlling actual detector bias, temperature and readout response. Examine reciprocal-density linearity only after justifying how the measured current maps to density. A successful fit supports the tested rate-law approximation; it does not uniquely identify the microscopic electron-hole recombination mechanism. Temperature and spectral sensitivity also vary in real PbS detectors and must be held or measured independently.[ref]
How does a PbS photoconductive detector work? by Shimadzu Scientific Instruments. temperature-dependence paragraph