A photoconductor transient with quadratic recombination is a carrier-density model in which the loss rate grows as the square of the mobile excess density. It explains how nonlinear loss can produce different turn-on and turn-off waveforms without assigning unrelated exponential lifetimes to the two directions. A photoconductor converts illumination-dependent electrical conductivity into a signal under external bias; the manufacturer's primary PbS description establishes that conversion mechanism.[ref]
How does a PbS photoconductive detector work? by Shimadzu Scientific Instruments. band excitation and external resistance readout
The quadratic loss law below is an explicitly stipulated model, not a fitted specification of that detector.Assume a spatially uniform excess density , generation rate , constant coefficient and an output proportional to at fixed mobility and bias:Density has units , has units and has units . A two-carrier recombination rate proportional to the product of equal electron and hole densities motivates a quadratic term in an appropriate regime. Background carriers, other loss processes, contacts and trapping can invalidate that specialization.
For constant , set and . Separating variables, or differentiating the following expression, gives the solution for any finite nonnegative initial density :For it reduces to . For it stays at equilibrium, and for it approaches the same equilibrium from above. In the dark, instead yieldsThe reciprocal-density plot is linear in this model, while a semilogarithmic decay plot is curved. The dark half-decay time is , so it depends on the starting illumination.
The small-signal lifetime under quadratic recombination is different again. Linearize about a nonzero background :A locally exponential small-signal response can therefore coexist with a strongly nonexponential large dark transient. One measured time constant cannot be substituted for all three quantities.
For an original example, stipulate and . Then , dark half-decay time , small-signal lifetime and small-signal half-power frequency about . Four times the background generation doubles equilibrium density and doubles that frequency in this model.
A proposed test measures small modulation about several backgrounds separately from complete turn-off traces, while controlling actual detector bias, temperature and readout response. Examine reciprocal-density linearity only after justifying how the measured current maps to density. A successful fit supports the tested rate-law approximation; it does not uniquely identify the microscopic electron-hole recombination mechanism. Temperature and spectral sensitivity also vary in real PbS detectors and must be held or measured independently.[ref]
How does a PbS photoconductive detector work? by Shimadzu Scientific Instruments. temperature-dependence paragraph
Articles by others on the same topic