Sampling correction in drug release testing is a material-balance adjustment for the substance removed, added or diluted when a receiver is sampled repeatedly. A measured concentration describes material still present at that moment; cumulative release also includes material withdrawn earlier. The distinction belongs to interpretation of an in vitro drug release test, whose transport and separation conditions are part of the measured system.[ref]
Let and be receiver volume and concentration just before the th withdrawal. At earlier observations , remove volume at concentration and add volume containing concentration . With initial receiver amount , conservation gives the cumulative amount contributed by the formulation:
The terms account for material remaining, material already removed and material supplied from elsewhere. The equation assumes that measured species is conserved and that the receiver is well mixed; adsorption or chemical transformation requires additional terms.
For initially blank medium, blank replacement and constant volume , this reduces to . If the aliquots are not replaced, the current volume instead satisfies .
Consider an illustrative receiver initially containing . Withdraw at without replacement. At the next observation, and , so the cumulative amount is . Incorrectly retaining the initial volume in the calculation would give . These chosen numbers illustrate accounting, not a formulation specification.
The sampling schedule and replacement protocol must therefore accompany a release curve. A decline in concentration after dilution does not imply that previously released material returned to the carrier. Similarly, a finite-difference drug release rate should be calculated from corrected cumulative amounts, with uncertainty and any shared measurement errors carried through the calculation.