An in vitro drug release test measures the amount or rate at which a substance leaves a formulation under specified laboratory conditions. The medium, temperature, agitation or flow, separation method and sampling schedule define the experiment. A measured release curve is therefore a property of a formulation–apparatus system under those conditions, rather than a direct measurement of the concentration in a person.
A typical design places the formulation in contact with a receiving medium, separates released material from the carrier and measures concentration at several times. The primary FDA technical account describes pressure-driven separation as an alternative to passive dialysis-membrane transport for complex formulations. It emphasizes that the separation step can itself limit the observed rate.[ref]
Adaptive Perfusion: A Novel In Vitro Drug Release Testing Method for Complex Drug Products by the FDA. Technology Summary, pressure-driven separation and limitations of conventional membrane methods
Repeated sampling requires a material balance. Suppose a well-mixed receiver has constant volume . At each observation a volume is removed and replaced with blank medium. If is the concentration immediately before the th withdrawal, the cumulative amount transferred into the receiver by time isThe first term is material still in the receiver; the sum is material removed at earlier observations. This sampling correction in drug release testing assumes matched replacement, a well-mixed receiver and no unmeasured loss or chemical transformation.
For illustrative , , and , the amount at the second observation is , not just . These chosen values demonstrate accounting; they are not a formulation specification or dosing instruction.
An informative experiment also accounts for material retained in the formulation and material lost to the apparatus. Adsorption, incomplete separation or degradation can change a measured concentration without the same change in carrier release. A control without the carrier and a recovery check address different parts of this problem. Comparing the measured curve under several justified apparatus conditions helps distinguish drug release kinetics from transport and analytical limitations.
Sampling correction in drug release testing is a material-balance adjustment for the substance removed, added or diluted when a receiver is sampled repeatedly. A measured concentration describes material still present at that moment; cumulative release also includes material withdrawn earlier. The distinction belongs to interpretation of an in vitro drug release test, whose transport and separation conditions are part of the measured system.[ref]
Adaptive Perfusion: A Novel In Vitro Drug Release Testing Method for Complex Drug Products by the FDA. Technology Summary, release measurement and apparatus transport
Let and be receiver volume and concentration just before the th withdrawal. At earlier observations , remove volume at concentration and add volume containing concentration . With initial receiver amount , conservation gives the cumulative amount contributed by the formulation:The terms account for material remaining, material already removed and material supplied from elsewhere. The equation assumes that measured species is conserved and that the receiver is well mixed; adsorption or chemical transformation requires additional terms.
For initially blank medium, blank replacement and constant volume , this reduces to . If the aliquots are not replaced, the current volume instead satisfies .
Consider an illustrative receiver initially containing . Withdraw at without replacement. At the next observation, and , so the cumulative amount is . Incorrectly retaining the initial volume in the calculation would give . These chosen numbers illustrate accounting, not a formulation specification.
The sampling schedule and replacement protocol must therefore accompany a release curve. A decline in concentration after dilution does not imply that previously released material returned to the carrier. Similarly, a finite-difference drug release rate should be calculated from corrected cumulative amounts, with uncertainty and any shared measurement errors carried through the calculation.
Articles by others on the same topic