Stefan Meyer, the mass-spectrometry researcher identified by the 2017 paper below, developed automated analysis of time-of-flight mass spectrometry data with collaborators. The paper lists his affiliation with Space Research and Planetary Sciences at the University of Bern. Its motivation includes handling large numbers of measurements for chemical-composition studies in space instrumentation.[ref]
Fully automatic and precise data analysis developed for time-of-flight mass spectrometry by Stefan Meyer and others. p.580, full byline, affiliation and Abstract
The analysis connects sampled waveforms to mass-related quantities rather than treating every peak position as a known species. Its instrument calibration model relates the flight time to mass-to-charge ratio throughwith instrument-dependent constants and . Identified isotopes provide reference points for fitting the time-to-mass scale.[ref]
Fully automatic and precise data analysis developed for time-of-flight mass spectrometry by Stefan Meyer and others. p.582, equation(2) and calibration paragraph
For an original sensitivity calculation, suppose the corrected flight time is and a timing displacement of is introduced while the fitted constants are held fixed. The inferred ratio changes byor . A0.1-percent timing change therefore produces about a0.2-percent change in this modeled mass-to-charge value. This is a consequence of the square law, not the paper's measured accuracy for an instrument.