Surrogate-matrix response equivalence means that a specified surrogate sample matrix supports an adequately comparable analytical response to the authentic sample matrix over the method's intended range. Equivalence is tied to an analyte, preparation, detector and decision criterion. It does not require the materials to have identical chemical composition.
ICH M 10 calls for assessment of differences in analytical matrix effects and extraction recovery between surrogate and authentic materials, and evaluation of response parallelism across the range.[ref] A useful experimental design processes authentic unspiked samples, authentic samples with known analyte increments and spiked surrogate material through the same procedure. The endogenous contribution in authentic material must be accounted for.
An original linear measurement model makes a one-point comparison's limitation visible. Let the surrogate response be , and the authentic response be , where is the analyte concentration in stipulated units. Both give response 10 at . At , however, authentic material gives 20, and the surrogate calibration reports . Agreement at the crossing point concealed both a slope difference and an offset.
Matched increments remove a constant endogenous offset: and in a linear range. Their ratio estimates , provided additions do not themselves change the sample matrix or preparation. Similar slopes alone still do not establish identical blank interference, stability or behavior outside that range.
Set a comparison tolerance appropriate to the analytical use and examine uncertainty as well as nominal agreement. A wide uncertainty interval containing equal response is not strong evidence of equivalence. Likewise, matching a ratio normalized by an internal standard may be useful for that method while concealing unequal absolute extraction recovery; state which response has been compared.