Bloch nuclear induction experiment 2026-10-18
In their 1946 letter, Felix Bloch, William Webster Hansen, and Martin Packard reported water at room temperature. A steady magnetic field along and an oscillating field along produced a signal in a pickup coil along , displayed on an oscilloscope. Some samples included paramagnetic additives to accelerate thermal equilibrium. Resonance fields at two frequencies gave the same field-to-frequency ratio, consistent with the proton value.[ref]
Nuclear Induction by Felix Bloch and others. p. 127, Nuclear Induction, entire letter; geometry, water, additives, readout and two-frequency check
The detailed paper describes glass water bulbs inside tuned receiving coils. Transmitter coils, shields and a conducting paddle controlled leakage. Field modulation at 60 Hz drove the horizontal oscilloscope sweep. A diode and audio amplification supplied the vertical trace. Vibration changed coupling: their largest difficulty. Coil tuning capacitors formed tank circuits. The actual receiver omitted the conceptual diagram's radiofrequency amplifier.[ref]
The Nuclear Induction Experiment by Felix Bloch and others. pp. 475–480, Figs. 1–4, APPARATUS and opening RESULTS; actual diode receiver versus conceptual amplifier
The orthogonal coil axes suppress direct transmitter pickup, but do not eliminate it. A field-dependent signal at the nuclear resonance must be distinguished from leakage and noise; merely detecting a radiofrequency voltage is insufficient. This is why the field sweep and the independently checked field-to-frequency ratio matter.