In vitro drug release test
ID: in-vitro-drug-release-test
An in vitro drug release test measures the amount or rate at which a substance leaves a formulation under specified laboratory conditions. The medium, temperature, agitation or flow, separation method and sampling schedule define the experiment. A measured release curve is therefore a property of a formulation–apparatus system under those conditions, rather than a direct measurement of the concentration in a person.
A typical design places the formulation in contact with a receiving medium, separates released material from the carrier and measures concentration at several times. The primary FDA technical account describes pressure-driven separation as an alternative to passive dialysis-membrane transport for complex formulations. It emphasizes that the separation step can itself limit the observed rate.[ref]
Adaptive Perfusion: A Novel In Vitro Drug Release Testing Method for Complex Drug Products by the FDA. Technology Summary, pressure-driven separation and limitations of conventional membrane methods
Repeated sampling requires a material balance. Suppose a well-mixed receiver has constant volume . At each observation a volume is removed and replaced with blank medium. If is the concentration immediately before the th withdrawal, the cumulative amount transferred into the receiver by time isThe first term is material still in the receiver; the sum is material removed at earlier observations. This sampling correction in drug release testing assumes matched replacement, a well-mixed receiver and no unmeasured loss or chemical transformation.
For illustrative , , and , the amount at the second observation is , not just . These chosen values demonstrate accounting; they are not a formulation specification or dosing instruction.
An informative experiment also accounts for material retained in the formulation and material lost to the apparatus. Adsorption, incomplete separation or degradation can change a measured concentration without the same change in carrier release. A control without the carrier and a recovery check address different parts of this problem. Comparing the measured curve under several justified apparatus conditions helps distinguish drug release kinetics from transport and analytical limitations.