A motorized micrometer uses a controlled motor to drive a precision screw and produce small linear motion. A typical design combines a lead screw, reduction gearing and a rotary encoder; the encoder may sense the motor shaft rather than the final load. The manufacturer's M-222–M-226 description makes this arrangement explicit and lists design resolution separately from minimum incremental motion, repeatability and backlash.[ref]
High Precision Actuators with Closed-Loop Feedback Motor by Physik Instrumente. M-221–M-226 description and technical-data table
For an original ideal drive, let screw pitch be per screw revolution, reduction motor revolutions per screw revolution and encoder count per motor revolution. The nominal output increment per count isA reducing lever with motion ratio 0.11 would give at its output per count. This is a kinematic increment, not demonstrated positioning accuracy. Compliance, backlash and instrumental drift can all exceed it.
A concrete experimental use advanced the specimen of an in-chamber ultramicrotome. Its 0.11 reduction transformed a quoted micrometer position measurement uncertainty of into a contribution at the specimen. The paper calculated nominal section thickness from the micrometer change and reduction ratio.[ref]
Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure by Winfried Denk and Heinz Horstmann. Materials and Methods, Microtome, specimen-advance mechanism and section-thickness calculation
That contribution is not a complete measured section-thickness uncertainty. An independent displacement check and reversals of motion help assess the actual response of an assembled mechanism. New to topics? Read the docs here!