Optical trap drag calibration
ID: optical-trap-drag-calibration
Translate the liquid past a bead held in a stationary harmonic optical trap. At steady mean position, take as the local fluid velocity relative to the fixed trap and positive downstream. Averaging the overdamped force balance givesFor the invented Optical trap Langevin calibration example, predicts , , and . Reversing reverses the mean shift but leaves the ideal thermal variance unchanged. A symmetric speed sweep tests the offset and the linear force range; the slope determines if position is calibrated.
Near a wall, is not the appropriate drag without a boundary correction; record bead height and direction of motion. Stage velocity must correspond to the local fluid velocity after transients have settled. Escape under increasing flow estimates a maximum holding force in that geometry, whereas small-displacement drag calibration estimates the local spring constant. They are different quantities. The atomic force microscopy similarly needs a calibrated spring constant to convert a measured deflection to force, but its cantilever is a physical spring rather than this optical field.