Phase-contrast magnetic resonance imaging
ID: phase-contrast-magnetic-resonance-imaging
Phase-contrast magnetic resonance imaging estimates motion from differences in the phase of nuclear magnetic resonance signals acquired with controlled magnetic field gradients. The encoding measures a component of velocity, with its sign set by the gradient and reconstruction convention. A 1982 paper described an early flow-sensitive phase-encoding approach.[ref]
For a constant-velocity teaching model, write position along the encoding direction as . If is the effective gradient, including refocusing signs, thenHere is the gyromagnetic ratio in radians per second per tesla, and is the gradient first moment. With balanced zeroth moments, two encodings giveThese equations are an explicit idealized derivation, not a claim that every measured voxel contains one uniform velocity.
Define the velocity encoding limit by . If it is , a corrected phase difference of corresponds to under the chosen sign convention. Phase wrapping modulo causes an ambiguity beyond the encoding range, related to aliasing.
A 1991 gated study measured differently timed cerebrospinal fluid motion in ventricular and subarachnoid compartments over the cardiac cycle.[ref]
Normal flow patterns of intracranial and spinal cerebrospinal fluid defined with phase-contrast cine MR imaging by D. R. Enzmann and others. Abstract; 16 flow-encoded images per cardiac cycle
Opposite velocities during equal parts of a cycle can average to zero despite substantial oscillatory motion. A volumetric flow rate additionally requires integrating the normal velocity across a defined cross-section; peak speed alone does not supply transported volume. New to topics? Read the docs here!