Phase-contrast magnetic resonance imaging estimates motion from differences in the phase of nuclear magnetic resonance signals acquired with controlled magnetic field gradients. The encoding measures a component of velocity, with its sign set by the gradient and reconstruction convention. A 1982 paper described an early flow-sensitive phase-encoding approach.[ref]
For a constant-velocity teaching model, write position along the encoding direction as . If is the effective gradient, including refocusing signs, thenHere is the gyromagnetic ratio in radians per second per tesla, and is the gradient first moment. With balanced zeroth moments, two encodings giveThese equations are an explicit idealized derivation, not a claim that every measured voxel contains one uniform velocity.
Define the velocity encoding limit by . If it is , a corrected phase difference of corresponds to under the chosen sign convention. Phase wrapping modulo causes an ambiguity beyond the encoding range, related to aliasing.
A 1991 gated study measured differently timed cerebrospinal fluid motion in ventricular and subarachnoid compartments over the cardiac cycle.[ref]
Normal flow patterns of intracranial and spinal cerebrospinal fluid defined with phase-contrast cine MR imaging by D. R. Enzmann and others. Abstract; 16 flow-encoded images per cardiac cycle
Opposite velocities during equal parts of a cycle can average to zero despite substantial oscillatory motion. A volumetric flow rate additionally requires integrating the normal velocity across a defined cross-section; peak speed alone does not supply transported volume.A bipolar gradient is a magnetic field gradient waveform with lobes of opposite sign. In nuclear magnetic resonance, such waveforms can control sensitivity to position and motion. Opposite signs alone do not guarantee equal signed areas, and an applied waveform must be distinguished from its effective encoding waveform when refocusing pulses change phase evolution.
For an original counterexample, apply for and then for . The gradient zeroth moment isThe waveform is bipolar but unbalanced.
If the two durations are instead equal to , and the effective lobes are contiguous, from time 0 to and from to , then but the gradient first moment is . Thus cancellation of stationary-position phase can coexist with sensitivity to constant velocity. The sign follows this stated ordering.
Bipolar pulses also appear outside flow imaging. A 1989 stimulated echo study addressed background gradients that distort diffusion measurements. Alternating pulses could cancel a cross term under specified conditions; tests used proton signals from homogeneous and granular samples.[ref]
Pulsed field gradient stimulated echo methods for improved NMR diffusion measurements in heterogeneous systems by R. M. Cotts and others. Primary abstract: background-gradient reduction, conditional cross-term cancellation and measurements
This is a sequence-specific result, not a promise that every bipolar pair removes every imaging error.The gradient zeroth moment is the signed time integral of an effective magnetic field gradient waveform along a specified encoding axis:Its units are tesla seconds per metre. Effective means that the sign convention accounts for any refocusing pulses in the nuclear magnetic resonance sequence; it is not necessarily the unsigned physical pulse area.
For a stationary spin at position , choose the phase convention . Then , where is the angular gyromagnetic ratio. A balanced waveform with cancels this position-dependent contribution within the model. It does not assert that every possible background contribution to phase vanishes.
As an original rectangular-pulse example, for has . Adding for cancels it. Equal areas, rather than equal pulse heights, produce the balance.
For a moving spin with , the remaining contribution can be , involving the gradient first moment. That separation explains how magnetic resonance velocity encoding can suppress sensitivity to stationary position while preserving motion information. Early phase-sensitive flow imaging supplied the experimental motivation for extracting such information from resonance phase.[ref]
A flow velocity zeugmatographic interlace for NMR imaging in humans by P. R. Moran. Primary abstract; displayed moment relations are explicit idealized derivations
The velocity encoding limit, commonly written , specifies the velocity scale represented by a phase difference of magnitude in a chosen phase-contrast magnetic resonance imaging encoding. With equal gradient zeroth moments and first-moment difference , an ideal constant-velocity model givesThe gyromagnetic ratio is angular, in radians per second per tesla, and is the difference of gradient first moments. A sign convention must be specified separately. A 1982 original paper described obtaining motion information through phase-sensitive nuclear magnetic resonance encoding.[ref]
A flow velocity zeugmatographic interlace for NMR imaging in humans by P. R. Moran. Abstract; phase encoding of flow information; displayed relations are the stated model
Choose the convention . If , then produces . But produces , which wraps to when phase is represented in . A reconstruction that ignores this aliasing would report . The limit describes an encoding ambiguity, not a physical maximum speed of the fluid.
For a fixed phase uncertainty in this simple model, propagation gives . Reducing improves velocity sensitivity under that assumption while narrowing the unambiguous range. For example, radians and give . Real measurements must also account for background phase, intravoxel variation and timing. Gated measurements of cerebrospinal fluid show why timing within the cardiac cycle matters even when net displacement is small.[ref]
The gradient first moment of an effective magnetic field gradient waveform is its time-weighted integral,Here effective includes the sign changes caused by refocusing pulses. Its units are when is in tesla per metre. The gradient zeroth moment is . For position with constant velocity, the accumulated nuclear magnetic resonance phase iswhere is the angular gyromagnetic ratio. Thus removes the dependence on initial position while can retain sensitivity to velocity, the basis of an idealized phase-contrast magnetic resonance imaging encoding.[ref]
A flow velocity zeugmatographic interlace for NMR imaging in humans by P. R. Moran. Original abstract; motion-sensitive phase encoding; equations here are an explicit constant-velocity derivation
For a teaching bipolar gradient, let the first lobe have signed area and time centroid , and the second have signed area and centroid . ThenEqual and opposite lobe areas cancel stationary-spin phase, but their time separation prevents cancellation of the constant-velocity phase. For and , the first moment is . These are illustrative waveform values, not measurements from the source.
Articles by others on the same topic