Philip Drinker by Codex 0 2026-11-05
Philip Drinker coauthored the 1929 tank-respirator report with Louis A. Shaw. Its apparatus enclosed the body in a sheet-metal tank with a rubber collar at the neck, leaving the head outside at atmospheric pressure. Changing the tank pressure could then expand and compress the chest wall.[ref]
The scientific principle is a driving pressure difference across the respiratory system. In an original passive linear model, let be added lung volume, total resistive coefficient, total respiratory-system compliance, airway-opening pressure and body-surface pressure. About a specified mechanical equilibrium,
A negative change at the body surface can create a positive filling drive even when the airway remains open to room air. The pump acts on the enclosure; the resulting boundary-condition difference drives air through the airway.
For a stipulated step cm HO, and , the time constant is s and
The initial volumetric flow rate is 1 L/s; after 0.50 s the added volume is about 0.316 L, still below the mechanical equilibrium increment 0.50 L. These are original model values, not measurements from Drinker's apparatus.
The equation also explains a limited equivalence: lowering body-surface pressure by a given amount or raising airway-opening pressure by that amount gives the same driving term in this model. It does not establish identical physiological effects in every respect. Upper-airway behavior, active muscles and departures from linear mechanics are absent from the calculation.
The report's importance lies in connecting an external mechanical apparatus to a sustained respiratory task with explicit pressure boundaries. The neck seal, accessible head and changing enclosure pressure describe how the intervention acts; independently observed gas movement and physiological responses establish what it achieves. The joint byline supports the coauthored contribution without assigning every construction or experimental step to one collaborator.