Bounding uncollected expiratory volume from terminal flow uses an explicit restriction on future volumetric flow rate to limit the air missed when a recorded expiration stops. A small final volumetric flow rate alone gives no such guarantee: an arbitrarily long, slowly decaying tail can have substantial remaining volume.
Let be the nonnegative future volumetric flow rate, with at the recording endpoint. Suppose an independently justified bound on its decay holds throughout the tail:
Differentiating shows that it cannot increase. Hence
This is an upper bound from a specified differential inequality, not a fitted exponential silently treated as established physiology.
For original values and s, the missed volume is at most 0.12 L. If the collected volume is L and the correctly captured first-second volume is L, then
If the terminal instrument reading has a deterministic error bound , use in the same argument. An uncertainty statement about the final reading does not itself establish the future decay assumption.
The assumption is essential. Exponential tails with the same terminal volumetric flow rate have missed volumes , unbounded as increases. This explains why the endpoint of a trace cannot by itself certify complete expiration. The primary spirometry statement separately specifies end-of-forced-expiration and measurement acceptability criteria.[ref] The inequality is an illustrative model bound, not an additional clinical stopping rule.

Articles by others on the same topic (0)

There are currently no matching articles.