Common-gain cancellation and flow-offset bias in expiratory ratios distinguishes an error that rescales both volumes equally from one that accumulates differently over time. Let true FEV1 be , true FVC be , and their ratio be . A shared positive volume gain cancels exactly: . The result requires the same gain for both quantities, not merely the same instrument name.
A constant error in measured volumetric flow rate has another effect. If the first-second interval lasts and the full recording lasts , integration gives
The ratio alone identifies neither nor separately; only their relative combination enters this model. Nor does the positive duration-dependent error necessarily raise the ratio.
The exact difference is
For small , its sensitivity coefficient at zero offset is . If average early volumetric flow rate exceeds average whole-record volumetric flow rate , a positive volumetric flow rate offset lowers the ratio. If the averages agree, that offset leaves the ratio unchanged in this model.
An original example uses L, L, s and s. The true ratio is 0.60. At , the zero-offset sensitivity is . An offset gives a first-order shift of ; the exact ratio is . A common 5% gain error without the offset would leave 0.60 unchanged.
Independent numerator and denominator gains would instead multiply by . Clipping, early termination and different timing conventions are additional measurement effects, so gain cancellation is not a general guarantee of ratio accuracy. The GUM requires the full measurement model and shared-input effects to be retained.[ref] The spirometry statement separately checks the acceptability of the constituent volumes.[ref]