Doppler range ambiguity means that a measured Doppler ultrasound signal does not uniquely identify the depth of the scatterers producing it. In continuous-wave Doppler echocardiography, simultaneous transmission and reception collect motion signals along the beam without a unique pulse travel time. A high-speed spectral component therefore does not, by itself, label an anatomical site.[ref]
Repeated pulses can introduce a different range ambiguity. Let the pulse period be , with pulse repetition frequency . If reception is associated only with its delay within a pulse period, the possible round-trip travel times include . The corresponding depths are
where is speed of sound. Identifying which transmitted pulse generated an echo requires additional information or a restriction on the possible depth.
For an original example, Hz gives microseconds. With m/s and observed within-period delay microseconds, possible depths are 7.7, 23.1 and 38.5 cm. Successive possibilities are separated by 15.4 cm. An anatomical restriction to depths below 15.4 cm would identify the first value; the echo timing alone would not.
Range ambiguity and aliasing of the Doppler ultrasound frequency are distinct. The first confuses position along the beam; the second confuses motion frequencies sampled at a given pulse rate. Increasing the pulse rate can improve the unaliased motion range while reducing unique depth coverage, producing the pulsed Doppler depth-velocity tradeoff. High-pulse-rate methods deliberately accept multiple possible sampling depths.[ref] Anatomical imaging and acquisition geometry supply constraints; a plausible waveform is not itself a unique range label.