End-expiratory hold 2026-11-05
An end-expiratory hold is an interruption of airflow at the end of exhalation during which airway pressure is observed while the respiratory system is occluded. In an appropriate passive, leak-free measurement it can assess total positive end-expiratory pressure, including intrinsic positive end-expiratory pressure. The set external pressure alone need not equal this measured total.[ref]
The mechanical reason is easiest to see in one communicating compartment. With airway resistance , let relate airflow to the difference between alveolar pressure and measured airway pressure. Once flow vanishes and pressure has equilibrated through the communicating airway, this model gives . Zero flow removes the resistive pressure drop; it does not mean that either pressure is zero.
For stipulated pressures, an external setting of and an equilibrated total of differ by . If plateau airway pressure is , respiratory driving pressure is , whereas subtracting the external setting would give 18. These constructed values illustrate reference selection, not patient settings.
Occlusion changes the boundary condition rather than directly opening every lung region to the sensor. A compartment isolated by airway closure cannot be assumed to equilibrate with the measured opening. Active muscular effort, a leak or a pressure trace that has not settled also defeats the simple passive-compartment inference. The observed airway-pressure plateau and the inferred regional alveolar pressure must therefore remain distinct.
Intrinsic positive end-expiratory pressure is the residual alveolar pressure above the externally imposed end-expiratory pressure, associated with incomplete lung emptying. In a communicating passive compartment,
The manufacturer's technical account makes this distinction and explains why a fitted whole-breath estimate can differ from the pressure obtained during an end-expiratory hold.[ref]
An original repeated-breath model isolates the timing mechanism. Let be excess volume above equilibrium with the outlet just before inspiration, and let each inspiration add tidal volume . During exhalation, assume a passive linear compartment with time constant . After expiratory time ,
The steady excess volume satisfies , hence
For stipulated L, respiratory-system compliance L/cm HO, s and s, , L and intrinsic pressure is about 4.66 cm HO. Giving this same ideal compartment two seconds to empty yields about 0.063 L and 1.25 cm HO. These are consequences of a model, not patient-specific timing instructions.
A nonzero external baseline does not by itself imply incomplete emptying. Conversely, setting an external baseline to zero does not prove that alveolar pressure has returned to zero. Measurements require attention to communication between regions, leakage and active muscle pressure; a single airway value need not reveal every trapped regional pressure. The recurrence explains why resistance, compliance and available time matter together rather than assigning the phenomenon to one pressure setting alone.
Positive end-expiratory pressure, PEEP, is airway pressure maintained above atmospheric pressure at the end of exhalation. It specifies an end-expiratory pressure reference, not a tidal volume or a pressure rise measured from the beginning to the end of inspiration. A set external pressure and the total pressure assessed during an appropriate end-expiratory hold need not be identical.[ref]
One reason is incomplete emptying. In an original passive single-compartment model, let be excess volume above equilibrium with the prescribed outlet pressure. With airway resistance and respiratory-system compliance ,
The remaining elastic pressure above the outlet is . Stipulated cm HO·s/L and L/cm HO give a time constant of 1 s. Starting with 0.50 L excess volume, after 0.5 s the model retains about 0.303 L, corresponding to 6.07 cm HO above the outlet pressure. These are constructed mechanics, not ventilator settings for a person.
If the prescribed outlet reference is 5 cm HO, the modeled retained elastic pressure raises the equilibrated total to about 11.07 cm HO. This illustrates intrinsic positive end-expiratory pressure: a pressure contribution that cannot be identified from the external setting alone. The result assumes a communicating passive compartment; active effort, leakage and heterogeneous time constants require further interpretation.
Respiratory driving pressure compares plateau airway pressure with total PEEP, rather than silently subtracting an external setting that may omit retained pressure. The manufacturer's primary measurement account states that distinction and the required passive, leak-free conditions.[ref] Naming the pressure reference is therefore part of defining the reported mechanical quantity.
Positive-pressure ventilation assists inhalation by raising pressure at the airway opening, driving gas into the lungs when that pressure exceeds alveolar pressure. It differs mechanically from spontaneous expansion of the thorax or negative-pressure ventilation, which lowers pressure around the lungs to produce the relevant distending difference. A ventilator's mask or tube supplies the applied airway connection.[ref]
The direction of flow still depends on a pressure difference, not on whether either pressure is above atmospheric. For a simple resistive path,
with positive directed inward and the airway resistance. In an original example, pressures 10 and 8 cm HO and cm HO·s/L give inward flow 0.5 L/s. Pressures 8 and 10 cm HO give outward flow 0.5 L/s. Both pairs are positive relative to atmosphere.
This explains why positive end-expiratory pressure can coexist with exhalation. The applied pressure can remain above atmospheric while elastic recoil produces a still higher alveolar pressure and drives gas outward. At equal opening and alveolar pressures, the same ideal resistive model gives zero flow.
The resulting tidal volume also depends on respiratory-system compliance, airway resistance, duration and any active respiratory muscle effort. The no-flow interval used to interpret plateau airway pressure removes one resistive contribution only under the required measurement conditions.[ref] Positive pressure is therefore a mode of applying mechanical assistance, not a guarantee of a particular inward flow, volume or gas-exchange result.