Respiratory driving pressure, , is the difference between plateau airway pressure and total end-expiratory pressure when measured under suitable passive conditions:The total baseline includes intrinsic positive end-expiratory pressure when present; substituting only the externally imposed baseline can overestimate this tidal pressure difference.[ref]
For the linear static model, respiratory-system compliance is . An original calculation with tidal volume 0.45 L, plateau pressure 23 cm HO and total baseline 8 cm HO gives cm HO and L/cm HO. If the external baseline alone were 5 cm HO, incorrectly subtracting it would give 18 cm HO and a different apparent compliance.
This pressure difference loads both the lung and chest wall. At static endpoints, transpulmonary pressure gives the partitionThus a stipulated airway increment of 12 cm HO corresponds to a lung increment of 4 when the pleural increment is 8, but to a lung increment of 10 when the pleural increment is 2. Equal airway driving pressures do not establish equal lung distension without information about the chest-wall contribution.
The measurement method matters as much as the subtraction. A dynamic estimate during flow can include resistance; an end-inspiratory hold and an end-expiratory occlusion address different endpoints. The primary pilot study explicitly compared static and dynamic ways of obtaining whole-respiratory-system compliance and driving pressure.[ref]
Static and Dynamic Measurements of Compliance and Driving Pressure: A Pilot Study by Pierre Tawfik and others. Introduction, Figure 1 and Study Procedures
These distinctions explain the quantity's mechanical meaning without turning an illustrative number into an individual ventilation recommendation. Articles by others on the same topic