Fluoroscopy produces a sequence of X-ray projection images displayed during an examination or procedure. An X-ray beam passes through the body and a detector records the transmitted pattern, allowing motion of anatomy, a catheter or an introduced contrast agent to be followed. This temporal information makes fluoroscopy useful in interventional radiology, where the position of an instrument must be checked as it moves.[ref]
A simple sampling example shows both the benefit and the limit. If a hypothetical system displays 15 distinct frames per second, successive frames are separated by seconds. An instrument moving at a constant travels about 2 mm between frames. The image sequence can therefore reveal movement that a single radiograph cannot, while still leaving motion between samples unobserved. These numbers describe an invented measurement, not a recommended operating setting.
Each projection also collapses depth: a needle and a target may overlap on the screen while remaining separated inside the body. Choosing another view or another imaging method can address that ambiguity. More frames alone do not restore the missing depth coordinate.
Because fluoroscopy uses ionizing X-rays, exposure duration and technique affect delivered radiation. Holding every other setting fixed while doubling the time doubles the accumulated exposure in a constant-rate model; actual procedures need their own dose measurements. The originating FDA description emphasizes that prolonged, complex image-guided procedures can deliver substantial doses.[ref]