Interferogram 2026-10-18
An interferogram records detected light intensity produced by interference as a controlled position, phase or delay is varied. In Fourier transform spectroscopy, the variable is the delay between two copies of the same field. Removing the constant background gives a correlation signal whose Fourier transform determines the source's spectral power density, subject to the measured delay window and instrumental response.
Interferogram apodization 2026-10-20
Multiplying a finite interferogram by a smooth window reduces sharp-edge artifacts. In the recovered radiation spectrum this convolves the true spectral density with the window's Fourier transform. Smoother windows generally lower sidelobes while broadening the main lobe, so a resolution figure must specify the window and criterion. Apodization does not recover unmeasured long-delay information.[ref]
Interpret the stated as the coherence-decay time, or the mean interval between fully phase-randomizing collisions in the simple impact model of pressure broadening. Then each line has correlation . The collisional phase relaxation and Lorentzian linewidth transform gives
The intrinsic spectrum is the sum of two such Lorentzian line shapes, still centred apart. With equal weights and balanced arms, its normalized visibility is
For the intensity normalization of the first source plot, the corresponding model interferogram is
The constant background remains ; only the oscillatory contrast decays. The beating zeros and recurrence spacing remain fixed, but successive visibility revivals shrink, reaching a factor at and at . Long-delay fringes disappear toward the background.
Figure 1.
Calculated pressure-broadened interferogram, decaying visibility revivals and Lorentzian doublet for a one-nanosecond coherence time
. Details.
The sketch chooses a carrier frequency for legibility; the envelope and linewidth follow from the inferred spacing and specified coherence time. A finite experimental delay window additionally convolves these intrinsic Lorentzians with the instrumental line profile. If were instead a population lifetime, the lifetime-limited field decay rate would be and the corresponding linewidth would be half the boxed value. The source does not define that time convention explicitly; pressure-collision coherence decay is the interpretation used here.
A non-polarizing beam splitter divides the incident spectrum into a fixed arm and an arm with a translating mirror. Recombining their returned fields gives a detector signal as a function of mirror position . The round-trip optical path difference is relative to the equal-arm position, so the delay is in air approximated as vacuum. Record both detector signal and calibrated position, subtract the constant background, and take a Fourier transform of the interferogram.
Figure 1.
Scanning Michelson spectrometer: mirror position sets delay, and a Fourier transform of the recorded interferogram recovers the spectrum
. Details.
For matched arm polarizations and balanced throughput, write the ideal signal as
where is the incident spectral power density and is a constant set by the beam splitter and detector. This is the cosine-transform form of Fourier transform spectroscopy, or equivalently the Wiener-Khinchin theorem applied to delayed self-interference.
The delay-range resolution in Fourier transform spectroscopy follows because recording a finite delay range multiplies the correlation by a window. Its transform broadens each spectral line by an instrumental width of order the reciprocal delay range. A mirror travel range creates a delay range of order , so
At fixed wavelength, absolute wavelength resolution therefore improves inversely with travel range. The numerical coefficient depends on whether is a one-sided or total travel, on the window or apodization, and on whether resolution means first zero, half width or another criterion. The proportionality is the requested convention-independent result. Position sampling must also be fine enough to sample the fastest fringe variation without aliasing.
In Fourier transform spectroscopy, a scanning Michelson interferometer measures intensity against optical path difference rather than selecting one narrow wavelength at a time. After subtraction of the constant background, an ideal stationary source gives an interferogram
A cosine Fourier transform recovers the spectral power density after instrumental calibration. A mirror displacement changes the round-trip path by ; the zero-path-difference burst aligns the scan, and a reference laser can calibrate the displacement.
Finite maximum path difference multiplies the interferogram by a window and convolves the recovered radiation spectrum with that window's Fourier transform. Thus is of order ; precise factors depend on scan range and resolution convention. Uniform path samples of separation need to avoid aliasing without bandpass information. interferogram apodization reduces sidelobes at the price of broader lines. The broad accepted optical throughput can be advantageous; the multiplex signal-to-noise advantage depends on detector noise and is not universal in photon-noise-limited measurements. Source drift during a scan, finite beam divergence, imperfect sampling and wave phase errors require control. The finite-window reasoning is the delay-range resolution in Fourier transform spectroscopy.[ref]
The two sodium lines are distinct optical frequencies, so their rapidly oscillating mutual cross terms average to zero at the detector. Assuming equal line strengths and negligible individual linewidths, add their separate interferograms. Write and . With the total mean output intensity,
For mirror displacement this becomes
This is spectral doublet from Michelson visibility revivals. Since , . Numerically,
The first interference visibility zero occurs at , with further zeros at odd multiples of that distance. Visibility is restored at integer multiples of . The signed envelope has period : alternate high-visibility packets have opposite fringe phase. Unequal line strengths leave a nonzero minimum visibility, and finite linewidths introduce a further decaying envelope; the wavelength data alone specify neither effect.
Figure 1.
Monochromatic Michelson fringes and the fast-fringe and visibility scales of a sodium doublet
. Details.
For mutually incoherent spectral components, the detected intensities add. If describes the mean output spectral intensity, the interferogram is
Subtract the constant background, scan the mirror with a calibrated displacement, and take the Fourier transform with respect to delay . This is Fourier transform spectroscopy. The transform recovers the spectrum subject to detector response, finite scan length and any window function. Sampling the displacement finely enough to resolve the shortest optical fringes avoids aliasing.
Using the usual reciprocal-maximum-delay resolution estimate, , and , gives
The mirror displacement is measured from zero optical path difference. This delay-range resolution in Fourier transform spectroscopy is a conventional estimate: a specified window and resolution definition introduce a numerical factor. For example, a symmetric rectangular delay window has its first instrumental zero at , which would give under that particular criterion. The paper does not specify a line-profile criterion; the standard estimate states its convention explicitly.
A Michelson interferometer uses a beam splitter to divide collimated input light between a fixed mirror and a moving mirror. Returning beams recombine at the beam splitter; one output reaches a photodetector. A compensator plate can equalize the glass traversed in the two arms. Moving one mirror by changes the optical path difference by .
The usual nominal Fourier transform spectroscopy specification is , where is the maximum optical path difference measured from equal arms. It gives
This is a nominal spectral resolution element convention.[ref] A symmetric rectangular scan from to instead has a sinc instrumental line shape with first-zero offset . Quoting that first-zero criterion would give cm; interferogram apodization and other linewidth criteria change the numerical factor. An exact mathematical minimum requires such a criterion.
In the measurement chain, the Michelson interferometer is the spectral analyzer or modulator, between the collection optics and the photodetector. It encodes the incoming radiation spectrum into an interferogram; detector electronics digitize that signal and a Fourier transform reconstructs the radiation spectrum. It does not itself turn radiant power into an electrical signal.
Use repeat-pass interferometric synthetic-aperture radar. An orbiting radar transmits coherent microwave pulses and records the amplitude and phase of the returning field. Form an interferogram from registered complex images acquired before and after the earthquake, remove the orbital contribution and the topographic phase predicted by a digital elevation model, and correct atmospheric and other long-wavelength phase delays. Where interferometric coherence is adequate, unwrap the residual phase to estimate line-of-sight displacement:
with sign set by the instrument's displacement convention. Each phase cycle represents of range change.
This maps a distributed earthquake displacement field day or night and through cloud, including areas inaccessible on foot. Roughly – ground sampling and a scene spanning many kilometres on both sides of the fault can resolve regional deformation; a narrow surface break benefits from finer pixels or complementary sub-metre optical surveys. Sensitivity to displacement can be millimetres to centimetres in coherent areas, but that does not equal spatial resolution. A single viewing direction measures one component; ascending and descending tracks and GNSS constrain more components. At the rupture itself, large gradients and changed ground cover can destroy interferometric coherence; radar pixel-offset tracking or before/after optical correlation can then recover metre-scale fault offsets.