Liquid scintillation counting measures radiation through light produced in a liquid scintillator containing the radioactive sample. Energy deposited in the medium leads to optical emission, which photomultipliers detect as pulses. The apparatus is a scintillation counter, but the measured pulse rate still needs a model connecting it to radioactive activity.[ref]
A Review of NIST Primary Activity Standards for 18F: 1982 to 2013 by Denis E. Bergeron and others. Section 2.1: scintillation medium, light conversion and photomultiplier detection.
In an idealized one-count-per-decay model with background subtracted and other corrections negligible,where is radioactive activity and is detection efficiency. A hypothetical net rate of with corresponds to . At the same activity, an efficiency of would produce only . A smaller count rate need not mean that fewer decays occur.
A motivating measurement example is L. L. Lucas's comparison of tritium-containing water standards. Liquid scintillation counting supplied massic activity ratios that could be compared with independently recorded mass-based dilutions. Agreement within stated measurement uncertainties supports consistency of the standards; it does not turn the counting apparatus into an uncertainty-free activity meter.[ref]
Massic Activity Ratios of the NBS/NIST Tritiated-Water Standards Issued Between 1954 and 1999 by L. L. Lucas. Abstract, Journal of Research of NIST 105 (2000), pp. 535–539; DOI 10.6028/jres.105.042.