Keithley Instruments is an instrument manufacturer associated with precision electrical measurement. Its Low Level Measurements Handbook, Seventh Edition by Keithley Instruments explains how current sources, voltmeters and connection arrangements determine what a low-level reading means, with concrete treatments of lead resistance and thermoelectric offset.[ref]
Low Level Measurements Handbook, Seventh Edition by Keithley Instruments. Chapter3, Measurements from Low Resistance Sources; §§3.3.1–3.3.2
A useful example is the handbook's comparison of two-wire resistance measurement with four-terminal sensing. In two-wire measurement, current and voltage readout share the leads, so their resistance contributes to the measured drop. Separate voltage-sensing connections can avoid including that lead drop when sensing current is negligible.[ref]
Low Level Measurements Handbook, Seventh Edition by Keithley Instruments. §3.3.1, Figures3-16–3-17, two-wire and four-wire arrangements
For an original ideal calculation, a resistor has two current leads of each. At , the total drop is , giving if all of it is attributed to the specimen. A separate voltmeter connected directly across the specimen, with negligible sensing current, instead measures and gives .
That result follows from the connection arrangement, rather than from the manufacturer's name. Finite input impedance, contact placement, thermoelectric offset and source uncertainty can still matter. The manufacturer's technical handbook is useful as an apparatus-level account of such limits; its example methods do not establish the accuracy of an unspecified instrument or a particular laboratory setup.
WIKA is a manufacturer of measuring instruments whose technical publications describe pressure measurements using elastic sensing elements. Its technical information distinguishes pressure-driven deformation from the linkage and pointer mechanism that turns it into a measurement indication.[ref]
Elastic element pressure gauges by WIKA. IN00.01,09/2010, p1, Description and labelled Bourdon-gauge mechanism
Its documentation is useful for connecting a physical measurement principle to the specifications of a complete instrument.A concrete specification also illustrates the difference between an error limit and measurement uncertainty. WIKA's June2025 US sheet for specified four- and six-inch gauges gives a base limit of of span and an additional temperature error of of span per away from the reference.[ref]
Bourdon Tube Pressure Gauges, Stainless Steel Series by WIKA. 23X.50,06/2025, p1, Accuracy class and Temperature error
For an original instrument example with span , the base bound is . A temperature departure contributes a further bound . Adding those magnitudes gives a conservative combined bound under this simplified interpretation; it is not a statistical root-sum-square uncertainty calculation. At a reading of , even the base bound is of the reading, despite being of span.
The example concerns the stated version and conditions, not a brand-wide accuracy claim. Use the actual model, range and pressure gauge temperature error specification together with instrument calibration evidence when interpreting a particular Bourdon tube pressure gauge.
Instrument maker associated with the Fortin barometer.