Fixed-cistern barometer Created 2026-10-31 Updated 2026-11-05
A fixed-cistern barometer is a mercury barometer whose reservoir geometry is left fixed during a reading. As mercury rises in the narrow tube, it is withdrawn from the reservoir and the reservoir level falls. The measured column-height change consequently exceeds the upper-surface displacement relative to the instrument frame.[ref]
For constant tube and reservoir areas and , conservation of mercury volume gives a reservoir drop when the tube surface rises by . Thus in this ideal geometry. A calibrated scale can incorporate that ratio; temperature changes require additional corrections. A Fortin barometer instead restores the lower surface to a fixed index before reading, illustrating two ways of controlling the same reference-level problem.
Hydrostatic equilibrium Created 2026-09-24 Updated 2026-11-05
Hydrostatic equilibrium is static balance between pressure gradients and a body force in a fluid. For upward coordinate and downward uniform gravity,The fluid can remain at rest because the higher pressure below a small fluid element supplies the upward force needed to balance its weight. At constant mass density, .
A connected pair of liquid columns or a mercury barometer makes the height-pressure relation measurable. In a compressible atmosphere the mass density varies with height, so an equation of state and thermal assumptions are needed as well. An isothermal atmosphere gives an exponential profile for an ideal gas, whereas constant density gives a linear profile. Hydrostatic equilibrium is a mechanical condition and does not by itself establish thermal equilibrium or stability against convection.
Mercury barometer Created 2026-10-31 Updated 2026-11-05
A mercury barometer measures atmospheric pressure by balancing it against a vertical Mercury column. A tube closed at its upper end connects below to a mercury reservoir open to the atmosphere. If the gas pressure above the column is , hydrostatic equilibrium gives , where is the vertical separation of the two mercury surfaces.
The height measures a pressure difference; interpreting it requires the actual mercury mass density and local gravitational acceleration. Residual gas above the column, capillarity and temperature-dependent scale length can shift the reading.[ref]
Mercury Barometers and Manometers by W. G. Brombacher and others. §3.1, principle of measurement; §§6.4 and 7, vacuum and error sources
The mercury-barometer temperature correction treats density and scale expansion explicitly. Distinguishing the two liquid levels is essential: an upper-surface position alone is not always the full column height.Observe each gas volume, temperature, pressure and moisture condition, and reduce both volumes to the same temperature, pressure and dry-gas basis. Equal scale readings alone do not establish equal amounts of gas.
Read the calibrated volume of each gas in a graduated tube or eudiometer, with the eye at the meniscus to avoid parallax. Allow the gas and the surrounding bath to reach thermal equilibrium, then read a thermometer close to the gas. Heating expands a fixed amount of gas, so a larger reading could reflect a higher temperature rather than more gas. The temperature used in a gas law must be an thermodynamic temperature.
Read the atmospheric pressure with a barometer. If a gas is confined over a liquid, also observe the difference between the inside and outside liquid levels. With the internal surface height above the external surface, the total internal pressure isHere is the confining density of the liquid and the gravitational acceleration; the sign reverses automatically when is negative. Equalizing the levels before taking the volume reading removes this hydrostatic pressure correction. A mercury barometer can itself require temperature and calibration corrections for accurate work.
If the gas is saturated with water vapor, subtract the vapor pressure of water at its measured temperature:The observed pressure includes both the test gas and water vapor; counting it all as test gas overestimates the amount. The saturation assumption needs time for equilibration. For a dry gas, no water vapor subtraction is made; another collecting liquid may require its own vapor pressure correction. A collecting liquid must also be chosen so that it neither appreciably dissolves nor reacts with the gas: water, for instance, is unsuitable for preserving the whole amount of freely soluble Ammonia.
For dilute gases obeying the ideal gas law, the volume at a common chosen isConsequently the corrected comparison isThis is also the ratio of amounts of substance in the ideal gas approximation. Check for leaks, trapped air, changing composition and incomplete transfer. At high pressure or near condensation, real gas behaviour may require compressibility factors; with , the inferred amount ratio contains the further factor .