NMR pulse calibration determines pulse settings that produce intended spin rotations under specified experimental conditions. For a resonant rectangular pulse, varying duration at fixed amplitude changes the NMR flip angle as , with the angular NMR nutation frequency. A pulse-acquire measurement follows how the resulting transverse magnetization changes.[ref]
Assume that each acquisition begins from the same magnetization, relaxation during the pulse is negligible, and the receiver phase reads the signed component after a positive- rotation. The ideal response is
A duration scan can locate an inversion null and give a nominal duration by halving the corresponding duration at unchanged amplitude.[ref]
For an original ideal data set with ,
The10microsecond pulse gives a transverse extremum, while the20microsecond null corresponds to inverted longitudinal magnetization. The null at zero duration is a different point of the same curve, so a single zero signal cannot identify the rotation. These values are constructed from the model, not recorded calibration data.
The phase convention matters: displaying only magnitude would hide the sign change beyond inversion. Reproducible initial conditions also matter; a sequence of differently timed pulses without consistent recovery would not test this one-pulse model. Sample-dependent RF loading and a distribution of effective field amplitudes can alter the measured response.
Calibration establishes a chosen pulse response in the tested setting. It does not by itself establish every offset response, cancel all NMR pulse errors, or validate a complete multi-pulse experiment. Recording the calibration conditions and the quantity being fitted makes later pulse-sequence comparisons interpretable.