Minimum-noise weights under baseline moment constraints choose a linear pulse estimator with the smallest variance among those that preserve a specified signal and remove specified backgrounds. Write the reading vector as , where is a known signal template, the columns of contain baseline shapes, and the error covariance matrix is positive definite. For , require
Columns in implement the moment conditions for baseline-drift cancellation. The noise cost of polynomial baseline cancellation depends on which feasible weights are chosen.
If the constraint columns are independent, Lagrange multipliers give
For any feasible perturbation , and , so the variance increases by . This proves the minimum and connects the estimator to generalized least squares. The covariance propagation is the full quadratic form, not a sum that discards correlations.[ref]
For an original one-bin pulse example, use five equally spaced times , a signal present only at the centre, and independent errors of variance . To remove just constant and linear backgrounds, the optimal dimensionless weights are . Their variance is , lower than for the centre and its two nearest references alone. Multiplying the amplitude estimate by a bin duration gives energy variance . Extra uncontaminated references can reduce noise when the cancelled background family stays the same.
The inverse requires genuinely independent constraints. Constraint redundancy in baseline-noise minimization calls for removing redundant columns and checking feasibility; a signal template inside the nuisance span cannot be preserved while that span is rejected. NIST's before/after pulse baselines provide the physical measurement context.[ref] The five-point design is an original idealization and assumes no pulse signal in its reference samples.
Moment conditions for baseline-drift cancellation specify which polynomial backgrounds a linear measurement removes. Let readings at known times enter the corrected pulse estimate as , with the integration/calibration units included in . A background contributes nothing whenever
Substitution gives this result directly. The zeroth condition removes a common offset; the first additionally removes linear drift. These conditions concern the actual weights and times, not only the number of background samples.
An original one-bin signal model measures a pulse at , with uncontaminated reference samples at and . Use
The weights sum to zero and have zero first time moment. Constant and linear backgrounds cancel exactly, but a quadratic background leaves bias . With chosen , and , the bias is . Here the pulse is stipulated to occupy the central bin; the formula is not a universal integral estimator for an arbitrary waveform.
Cancellation has a noise cost of polynomial baseline cancellation. For independent equal-variance reading errors , the chosen estimate has variance , even though its constant and linear bias terms vanish. General correlations use instead.[ref]
NIST's historical pulse-energy apparatus used before/after baselines for background and thermal-drift compensation.[ref] The polynomial test above is an original general analysis. Pulse contamination in a reference, uncertain timing or drift outside the assumed polynomial family can leave a residual even when the nominal algebraic conditions hold.