An angle-resolved photoemission spectroscopy intensity map locates occupied electronic band structure through its bright dispersing ridges. For a nearly isotropic graphene cone,
Fit straight lines to the reasonably linear portions of the ridges, avoiding the broad central crossing, and extrapolate them to locate the Dirac point energy . The PDF places the crossing near and plotted energy roughly –. On a lower branch, a representative change from approximately towards gives a slope near . Allowing for the ridge width gives about –, hence
Here .
To determine the carrier Fermi energy, identify the Fermi level from a calibrated photoemission cutoff, then measure its separation from the Dirac point:
The separation of the two Fermi crossings gives for a cut through the cone centre. This determines the doping energy relative to the Dirac point; the absolute plotted energy origin is not itself a Fermi energy.
The reproduced panel has no marked Fermi level or stated energy-zero calibration. Its upper ridges end near , but that alone does not prove that the endpoint is a Fermi cutoff. If it is the intended cutoff, the estimate is – and –. If instead the plotted numbers are binding energies measured from an unstated Fermi zero, the inferred separation would be roughly –. The supplied figure securely supports the slope estimate, but it does not distinguish these energy-reference conventions; an unconditional numerical Fermi energy requires the missing calibration.