Air-displacement pipette 2026-11-05
An air-displacement pipette uses a piston to change the pressure and volume of an air cushion above a liquid sample. On piston withdrawal, reduced pressure allows ambient pressure to drive liquid into the tip; subsequent piston motion expels it. The piston itself remains separated from the sample.[ref]
Gilson Guide to Pipetting. second edition, pp. 8–11, air-cushion mechanism and aspiration/dispensing sequence
The air cushion explains why a mechanical setting alone does not establish the delivered volume. Its response can depend on temperature and sample properties, including volatility and density. A positive-displacement pipette uses direct piston contact instead of that intervening gas volume.[ref]
Gilson Guide to Pipetting. pp. 9–12, contrast between mechanisms and influence of sample properties
For an original repeated-delivery example, compare sets of settings with measured mean deliveries of and . Knowing those means alone does not identify which set has less scatter. Both the mean error and a dispersion measure such as standard deviation are needed to describe performance. A microlitre is an exact unit relationship; an individual micropipette delivery is an experimental measurement.
Micropipette 2026-11-05
A laboratory micropipette is a pipette for transferring small measured quantities of liquid, commonly expressed in microlitres. In piston-operated models, the displacement mechanism matters. An air-displacement pipette leaves an air cushion between piston and sample, whereas a positive-displacement pipette places the piston directly against the liquid.[ref]
Weighing a delivered sample provides a concrete check. In a simplified calculation that neglects air buoyancy and evaporation, liquid volume follows from mass and density :For an assumed , a delivered mass of corresponds to . Relative to a nominal setting, the deviation is . These are illustrative values, not calibration data.
A repeated series can distinguish a displaced mean from wide scatter: measurement accuracy and standard deviation describe different aspects of performance. Actual gravimetric checks account for temperature-dependent water density, air buoyancy and evaporation losses. The manufacturer's method uses a conversion factor and an evaporation correction, showing why a displayed volume alone does not establish the quantity delivered.[ref]
Gilson Guide to Pipetting. pp. 46–49, systematic/random error, gravimetric conversion and repeated weighings
Positive-displacement pipette 2026-11-05
A positive-displacement pipette transfers liquid with a piston that contacts the sample directly, avoiding the intermediate air cushion of an air-displacement pipette. In the handheld designs described by the manufacturer, the capillary and piston form a disposable assembly. The mechanism can reduce effects associated with the expansion or contraction of trapped air.[ref]
Gilson Guide to Pipetting. second edition, pp. 8–11, positive-displacement construction and operation
This distinction is useful when handling liquids whose viscosity, density or volatility differs substantially from water. The instrument still measures a volume: a selected does not imply a fixed mass for every liquid. For an illustrative density of , the corresponding mass is by .
A gravimetric check therefore needs the appropriate density conversion. Direct displacement also does not make every delivery perfect: instrument condition, the fitted assembly and the test procedure remain relevant. Repeated measurements distinguish a systematic displacement of the mean from scatter in individual deliveries.[ref]
Gilson Guide to Pipetting. pp. 46–49, performance assessment and gravimetric procedure