A preset tare device subtracts a preset tare value from a measured gross or net indication and displays the calculated result.[ref] A stored container value can therefore be reused without weighing each empty container individually.
Suppose a container and its contents have measured gross mass and the entered preset tare value is . Subtraction gives a calculated net mass of . If this particular container actually has tare mass , however, its contents have mass . Even an otherwise exact gross measurement would then overestimate the contents by .
The error follows without any instrument-specific assumption:
A mismatch in the entered container value passes directly into the result. Reusing one value for nominally identical containers is therefore an assumption about their tare masses, rather than a new measurement of each container.
The starting indication also matters. If an ordinary tare device has already removed a container contribution, the displayed is already net for that reference. Subtracting the same again would give . A deliberate second subtraction can represent another distinct component, but applying the same container correction twice is a different calculation from removing its tare mass once.

Articles by others on the same topic (0)

There are currently no matching articles.