A tare device establishes a zero indication on a weighing scale while a load remains on its load receptor. An additive tare device preserves the weighing range for net loads; a subtractive tare device reduces that range.[ref]
For example, place an empty container on the load receptor and operate the tare device. The display reads zero even though the scale still supports the container. Adding of material gives gross mass and, ideally, an indicated net mass of . Removing both material and container without clearing the tare would ideally give . This negative indication is consistent with the changed reference.
The cancellation can be seen in a linear model. If the unrounded indication before tare is , and the supported container has tare mass , the indicated result after adding net mass is
The additive offset and the container contribution cancel. A response scale factor remains, so taring alone does not establish calibration across the load range.
The container's physical load remains on the scale after taring. Consequently the available maximum weighing capacity depends on the kind of tare device, even when two instruments show the same zero. A preset tare device performs a related numerical subtraction using an entered or recalled value, rather than necessarily measuring this particular empty container.
An additive tare device is a tare device that can establish a zero indication with a load present without reducing the weighing range available for net loads.[ref] The supported container still contributes tare mass; the instrument provides a separate tare allowance rather than obtaining extra capacity simply by changing a displayed number.
Consider a deliberately specified model with maximum weighing capacity for net loads and a maximum additive tare allowance . Assume these ranges are available together for the stated configuration:
A container with tare mass can then accompany net mass , giving gross mass . The model's full combined boundary is when both allowances are used. These numbers are chosen examples, not specifications for a particular scale.
By comparison, a subtractive tare device with a gross-load budget would leave only for net mass after the same container is tared. Both displays can initially read zero; their available net-load capacities differ.
The OIML definition of maximum weighing capacity explicitly omits additive tare capacity, while its maximum-tare-effect terminology records the tare device's capacity separately.[ref] A complete load description therefore identifies the container's tare mass, the desired net mass and the instrument's declared tare allowance. The load receptor continues to carry their combined physical load.
A subtractive tare device is a tare device whose operation reduces the weighing range available for net loads.[ref] Its zero display accounts for a load already present, while that load continues to occupy part of the instrument's capacity.
Consider a chosen model with maximum weighing capacity and no additive tare capacity. If a container of tare mass is supported on the load receptor, the admissible net mass in this model satisfies
With , the largest net load in this capacity budget is . Adding after taring would produce gross mass , even though the desired net indication is below .
Changing to a container would leave available for net mass in the same model. This is a physical consequence of reducing the supported tare mass, rather than a benefit obtainable by repeatedly resetting the display.
An additive tare device is defined by retaining the net-load weighing range. Therefore the equation above is a model for subtractive tare and a shared gross-load bound, rather than a universal rule that every scale's printed includes every possible tare allowance. The chosen container, allowed tare range and instrument capacity must describe the same load configuration.