Mass analyzer 2026-11-05
A mass analyzer is the part of a mass spectrometer that makes ions distinguishable or selectively transmits them according to mass-to-charge ratio. The measured coordinate can be position, arrival time or transmission under a specified field setting. A separate mass spectrometer detector records the resulting signal.
A quadrupole mass analyzer selects trajectories using combined static and time-varying electric fields: chosen mass-to-charge values remain on stable paths while others are lost from the transmitted beam.[ref] A time-of-flight mass spectrometer instead encodes the ratio in arrival time after acceleration. Different physical designs can therefore provide the same kind of ionic-mass information through different observables.
An original magnetic example exposes an important assumption. For speed perpendicular to a uniform magnetic field , balancing the Lorentz force and circular acceleration gives
If the incoming ions have the same selected speed and charge, mass values in the ratio give bending radii in that ratio, about . But if ions instead start negligibly slowly and gain energy from the same electric potential difference , substituting gives . Their radius ratio is then . The initial-condition choice changes the mapping.
Analysis requires this mapping and its assumptions, not just separated peaks. Mass resolving power describes the ability to distinguish nearby peak coordinates under a stated width criterion; accurate assignment of those coordinates additionally needs instrument calibration and the correct charge-state interpretation.
Quadrupole mass analyzer 2026-11-05
A quadrupole mass analyzer is a mass analyzer that selects ions through stability of motion in combined static and alternating quadrupole electric fields. Selected mass-to-charge ratios follow bounded transverse trajectories through the analyzer; others grow unstable and fail to reach its exit.[ref]
For an idealized model, define the electric potential
Here is a geometric scale and follow this stated potential convention. An ion with electric charge and mass then obeys
At a given instant the field focuses one transverse direction and defocuses the other. Alternation can nevertheless yield bounded motion in both directions. Rescaling time reduces these equations to forms of the Mathieu equation; a quadrupole stability diagram summarizes the relevant bounded-motion regions.
The equations also show why the filter selects a ratio. Doubling both and leaves the ideal equations unchanged. With fixed geometry, the dimensionless coefficients scale as and . Doubling the drive angular frequency while keeping the potentials fixed divides both coefficients by four; multiplying both potentials by four restores them in this model.
A stable mathematical trajectory is only part of instrument performance. Initial ion positions and velocities, finite flight length, field imperfections and detector response also influence transmission and measured mass resolving power. The ideal potential is an explicit model, not a guarantee about a particular instrument.
A quadrupole stability diagram plots the parameter combinations for which an ideal quadrupole mass analyzer gives bounded transverse ion motion. The useful region is the overlap of stability in both transverse directions: confinement along only one direction still allows loss along the other. The physical selection mechanism is stability in combined static and high-frequency quadrupole electric fields.[ref]
Using the potential convention stated in quadrupole mass analyzer, introduce dimensionless time and parameters
Here is electric charge, is mass, describes geometry, is drive angular frequency, and specify the static and alternating potentials. The transverse equations become
These are Mathieu equations. In the convention , the first equation has ; quoting a diagram requires checking this sign and the potential's numerical convention.
A useful limiting test is , . The equation oscillates, but the equation has exponentially growing solutions. Static focusing alone therefore fails to provide a stable overlap. Periodic focusing changes this conclusion; its stability regions cannot be read off from the force at one instant.
At fixed positive , changing mass-to-charge ratio moves a point along a ray with . Doubling halves both coordinates. Doubling both applied potentials moves the same ion twice as far along that ray. These scaling relations explain how scanning the potentials selects successive mass-to-charge ratios. A stable region is a property of the ideal differential equations; finite rod length, entrance conditions and field imperfections still determine actual transmission and mass resolving power.