Analytical ultracentrifugation Created 2026-10-20 Updated 2026-10-31
Analytical ultracentrifugation observes concentration profiles during rotation, rather than only recovering a separated pellet. Sector-shaped sample cells rotate past optical detectors. Absorbance scans measure attenuation at selected radii. A Rayleigh interferometer compares a sample sector with a solvent reference, converting concentration-dependent refractive-index changes into fringe shifts. Repeating scans follows a sedimentation boundary or the approach to a stationary distribution.[ref]
For a dilute solute of buoyant mass and translational friction , radial force balance gives , where is the sedimentation coefficient. Diffusion supplies . A velocity experiment resolves boundary motion and broadening; a sedimentation equilibrium experiment instead measures . This follows from the Boltzmann distribution in centrifugal potential . The two approaches separate frictional size information from buoyant mass information; shape, hydration and solvent properties must be specified to infer a unique geometric dimension.
Rayleigh interferometer Created 2026-10-20 Updated 2026-11-05
A Rayleigh interferometer compares optical paths through a sample and a reference. A path-length difference shifts the interference phase by . In analytical ultracentrifugation, sample and reference sectors produce concentration-dependent fringe shifts, allowing radial concentration scans without requiring the solute to absorb at the probe wavelength.[ref]