Sampling skew after matched ratiometric filters is the timing difference between the signal and reference samples in a ratiometric measurement, after their analog paths have the same response. Shared wiring and matched filters cancel excitation at the same effective time; they do not make two different times interchangeable.
For constant sensor and reference resistances, let the common filtered excitation be , where is the common impulse response and denotes convolution. With consistent initial conditions, the two filtered electric potential differences are and . If the sensor is sampled at but the reference at , the measured ratio is
Both denominator factors must be nonzero. The sampling instants here mean the effective acquisition times; conversion-result arrival times need not be those instants. Ratiometric cancellation with matched signal and reference filters treats the equal-time cancellation, supported by the primary circuit report's matching requirement.[ref]
For positive differentiable and small skew, expanding the logarithm gives
A later reference sample during a rising excitation therefore makes the inferred resistance low. The relevant rate is the fractional slope , not the absolute slope alone. A timing-skew bound for a ratiometric measurement can express a worst-case requirement using a bound on this logarithmic slope.
For an original local example, suppose over the acquisition window, with and . Then the exact scale error is . The small-skew estimate is . Doubling the common amplitude leaves both errors unchanged; reversing the sample order changes their sign to first order. This is a stipulated short-window excitation model, not an instrument-performance claim.
A focused bench check replaces the sensor with a stable resistor, imposes a controlled excitation change, and compares simultaneous acquisition with deliberately delayed reference sampling. Recording the excitation on an oscilloscope supplies its fractional slope. Repeating on a settled plateau separates timing error from reference-resistor error in ratiometric resistance measurement and other steady scale errors.
A timing-skew bound for a ratiometric measurement limits the resistance-ratio error caused by sampling the signal and reference at different times. It converts a bound on the fractional rate of change of the shared excitation into an acquisition-timing requirement. This is useful even when the two analog filters are exactly matched.
Use the model in sampling skew after matched ratiometric filters: the common filtered excitation is , the true resistance ratio is , and the reference is sampled after the signal. Then . Suppose is continuously differentiable and its logarithmic derivative obeys throughout the entire interval between the samples. The fundamental theorem of calculus gives
The same proof works for either sign of . Writing , exponentiation yields the exact interval
This is a deterministic bound under the stated slope assumption, rather than an expansion truncated at first order.
For an original timing budget, suppose and require at most fractional resistance error from skew alone. It suffices to require
The slightly smaller-than- value accounts for the exponential upper bound. An excitation proportional to reaches one side of the logarithmic bound exactly, so the bound is attainable in this model.
A sufficient way to establish is to verify and , giving . A filter cutoff by itself supplies neither a positive excitation floor nor this fractional-slope bound. Near an excitation zero, a small absolute change can produce a large resistance-ratio error. The primary filter-design report establishes the need to match signal and reference dynamics and to allow filter settling; the inequality above adds an original unequal-time error budget.[ref]
A proposed bench check uses a stable sensor substitute and reference resistor, records the filtered excitation with an oscilloscope, and deliberately changes the acquisition delay. Compare each observed ratio with the interval predicted from a verified excitation floor and slope bound. This isolates timing effects from a fixed reference-resistor error in ratiometric resistance measurement.