LiftMode 2026-10-20
A two-pass scan method recording topography first, then retracing it at an offset height to measure a longer-range interaction with reduced contact contributions. It is a Bruker mode name.[ref]
LiftMode is a two-pass measurement separating a measured topographic path from a longer-range interaction measurement. The first pass records height with ordinary topographic feedback. On the second pass the tip retraces that line at a chosen vertical offset above it, reducing contact and short-range contributions while detecting electrostatic or magnetic interactions.
The lift height is relative to the recorded path, not a single constant laboratory height over a corrugated sample. Its usefulness depends on first-pass accuracy and whether the relevant long-range signal remains detectable at that separation.[ref]
Use Kelvin probe force microscopy to quantify a local contact-potential difference. A conducting tip and sample form a position-dependent capacitor. Apply a voltage with adjustable direct and small alternating components,
The electrostatic force is proportional to , so its component at is proportional to
up to the chosen force-direction convention. A lock-in amplifier detects this component and a second controller adjusts to null it. At the null, . The compensation voltage becomes the potential image.[ref]
A lift pass can reduce direct topographic interactions. Calibration on a reference with a known work function is required for an absolute work-function interpretation, and the sign convention must state which electrode is biased. On a semiconductor the result can include surface band bending, trapped charge and surface chemistry, so it need not equal a simple map of bulk electrochemical potential.