Thermodilution estimates blood flow from the temperature change produced by introducing fluid at a different temperature. It is a form of indicator dilution: the transported signal is a thermal deficit or excess rather than a dye concentration. An upstream catheter introduces the fluid and downstream temperature measurement records its dilution in flowing blood.
Two arrangements illustrate the principle. A finite injection produces a temperature-versus-time curve; a continuous infusion can produce an approximately steady downstream temperature change. A primary coronary-flow study used room-temperature saline, a 2.8-F infusion catheter and a 0.014-inch pressure/temperature guidewire, testing the method in five instrumented dogs and 35 patients.[ref] The quantities observed were infusion rate, infused-fluid temperature at the catheter tip and distal mixed-blood temperature.
For an idealized finite-injection model, let be the injected thermal deficit in joules, the volumetric heat capacity of the moving fluid, and the downstream temperature deficit. Assume steady volume blood flow , complete passage of the signal through the detector, representative mixing and negligible heat transfer to other structures. Conservation of energy then gives
The time integral, rather than the maximum temperature change, determines this model's flow estimate. With stipulated J, J/(mL K) and curve area K s, the inferred blood flow is mL/s, or mL/min. These values illustrate the balance, not a clinical protocol.
Figure 1.
Equal thermal-indicator area with different peak temperatures
. Synthetic deficits , with K s and or s. Both integrals over all time equal ; the plotted window ends at 16 s. Different peak heights therefore give the same ideal flow at fixed thermal input. Details.
This ideal calculation exposes measurement assumptions. If only a fraction of the thermal signal reaches the detector, the observed curve area is reduced by that factor and the uncorrected estimate becomes . For example, would cause a overestimate in this model. Coronary thermodilution mixing assumptions similarly determine whether a local sensor reading represents the transported thermal signal. Accurate temperature measurement alone cannot establish those conditions.
Coronary thermodilution mixing assumptions are the conditions that make a downstream temperature reading representative of the mixture used to infer coronary blood flow. They include adequate mixing of infused fluid and blood, appropriate volumetric heat capacity values, known infusion rate and temperature, and sufficiently small unmodeled heat transfer between the fluid and surrounding structures.
The apparatus places these assumptions in a concrete setting. In a primary continuous-infusion study, a catheter supplied room-temperature saline while a pressure/temperature guidewire measured distal temperature; the infused-fluid temperature was measured at the catheter tip.[ref] Temperature at the tip matters because fluid can exchange heat before entering the measured blood flow.
An original two-stream model makes the bookkeeping explicit. Let native blood flow be , infusion rate , baseline blood temperature , infused-fluid temperature and mixed temperature . Assume equal constant volumetric heat capacity, no external heat transfer and steady complete mixing. Conservation of energy gives
and hence
Native flow and total downstream mixture flow have different numerators. With constructed mL/min, , and , the model gives mL/min and mL/min. This equal-capacity balance is an explanatory model, not the instrument's full calibration algorithm.
Incomplete mixing changes what the sensor observes. If a sensor sampled an unusually cold filament and read for the same other inputs, the formula would return about mL/min instead of 80. Repeating that local reading precisely would not correct its lack of representativeness. More generally, a mixed temperature is weighted by transported heat capacity, rather than being an unweighted spatial mean.
Finally, becomes sensitive to small temperature errors when is small. At fixed ,
For the constructed example this is mL/(min K). The derivative describes local sensitivity within the model; it does not estimate errors from incomplete mixing or heat loss. Those require separate physical evidence.