Volumetric analysis determines an amount of substance from the measured volume of a reagent of known strength that reacts with it in a known proportion. A measured aliquot is placed in a suitable vessel, the standard solution is delivered from a burette, and the volume required to reach the chosen titration endpoint is read. A pipette measures the aliquot; a volumetric flask prepares or dilutes the standard solution.
For a reaction , if the concentration of the titrant is and the corrected volume delivered is , then
With volume in litres and concentration in moles per litre, the answer is an amount in moles. The corresponding mass is , using the appropriate molar mass. For an unknown solution of aliquot volume ,
These are modern symbols for the same proportional reasoning that a nineteenth-century analyst could express with equivalent weights and a specified standard solution.
A suitable reaction is sufficiently rapid, complete and selective, with a recognisable endpoint. In an acid-base titration, an indicator can change colour near neutralization; other reactions use precipitation, redox or a colour intrinsic to a reagent. The observed titration endpoint must represent the required equivalence point closely enough. The reagent's nominal preparation alone is insufficient if its concentration changes in storage; standardisation of a solution, a reagent blank where needed, and correct reading of the meniscus control systematic error. A titration measures the intended analyte only insofar as interfering substances have been excluded or accounted for.