Optical pulse temporal contrast
ID: optical-pulse-temporal-contrast
Optical pulse temporal contrast compares a pulse's peak with weaker light at another time. For a radiant power trace with peak , explicitly choose either the peak-to-local ratio or its inverse . The first is at least one where the local power is positive; the second lies between zero and one. With a fixed transverse beam shape, the same ratio can be formed from light intensity. If the beam shape changes with time, specify the spatial sampling instead of identifying local intensity and total power.
A plateau at of the peak has and . Expressing this power ratio as optical contrast in decibels givesFor an amplitude ratio whose power ratio is , the same definition gives . Contrast also needs a delay or time range: an optical prepulse, an optical postpulse and a long optical pulse pedestal can have the same relative height but different positions and laser pulse energies. The pedestal energy fraction additionally requires integration over the contribution's duration.
A useful nonlinear measurement is third-order intensity autocorrelation. In an ideal instantaneous model, an input light intensity generates a second-harmonic gate proportional to . Sum-frequency generation with a delayed fundamental replica gives a recorded signal proportional toThe instrument measures this cross-correlation, rather than simply reading . Finite gate width, phase matching, spatial overlap and detector response enter the interpretation.
The1998/99 instrument of J. Collier and colleagues used second-harmonic generation, a Wollaston prism and an optical biprism to form crossing fundamental and harmonic beams. A second crystal generated the sum-frequency signal, imaged through optical filters onto a charge-coupled device. Position encoded delay. An inserted glass etalon produced a delayed replica near of the main signal, recorded on one side of the correlation peak. Figure4 combines window settings shifted with etalons.[ref]
A Single Shot 3rd Order Cross-Correlator for Pulse Contrast and Pulse Shape Measurements by J. Collier and others. Printedpp194–195, Description/Figure2 and Results/Figure4
An independent ideal calculation explains both the asymmetry and a possible third-order pulse correlation ghost. Let represent two identical-shape pulses with well-separated supports and correlation lobes, with the second weaker and delayed. Squaring introduces no overlap term. If , expansion givesThe delayed replica produces a lobe of relative order , while the opposite-delay ghost has order . For , the scales are about and respectively. These values are predictions of the stated ideal model. Assigning every small correlation feature directly to a separate input pulse would miss this distinction.
A nondetection below the measurement floor establishes a bound under a detection model, rather than demonstrating that no light exists there. Finite dynamic range, detector baseline subtraction, delay coverage and resolution should accompany a reported contrast trace. The Sequoia manufacturer likewise distinguishes optical temporal resolution from the scan step; making that step smaller does not alone narrow the optical response.[ref]
Sequoia temporal contrast measurement by Amplitude. January2024 sheet, PDFp2, Specifications