Resistance thermometer immersion test

ID: resistance-thermometer-immersion-test

A probe's sheath and leads carry heat between the sensing region and the surroundings through thermal conduction. In a stable, sufficiently uniform heat bath, compare the indicated temperature as insertion depth increases, allowing settling each time. A plateau supports adequate immersion for that apparatus and required uncertainty; it does not prove that every depth or medium works. Keep the sensing element and reference thermometer at comparable depths, and check bath gradients independently, so moving the probe does not merely sample a different temperature. The documented platinum resistance thermometer comparison calibration includes these controls.