A gigaohm seal is the high electrical resistance contact between the glass rim and cell membrane, also called a gigaseal. It suppresses a bypass route around the membrane patch; it does not make the membrane patch itself impermeable. The 1981 methods used pulled, coated and heat-polished pipettes, a chlorided silver electrode and a pressure connection. Clean glass and suction supported tight seals; coating reduced pipette-to-bath capacitance and recording noise, rather than being required for sealing.[ref]
Seal leak current bypasses the membrane patch through an imperfect seal. For an ideal ohmic seal to a zero-voltage bath, , positive from pipette to bath. At and its magnitude is , far larger than the synthetic single-channel step. A stable baseline can be subtracted, but changing leakage can mimic a biological response. Intrinsic membrane leak through ion channels is distinct from this bypass: interpreting both as the same electrical conductance obscures whether the biological cell or the seal changed.