Air-displacement pipette 2026-11-05
An air-displacement pipette uses a piston to change the pressure and volume of an air cushion above a liquid sample. On piston withdrawal, reduced pressure allows ambient pressure to drive liquid into the tip; subsequent piston motion expels it. The piston itself remains separated from the sample.[ref]
The air cushion explains why a mechanical setting alone does not establish the delivered volume. Its response can depend on temperature and sample properties, including volatility and density. A positive-displacement pipette uses direct piston contact instead of that intervening gas volume.[ref]
For an original repeated-delivery example, compare sets of settings with measured mean deliveries of and . Knowing those means alone does not identify which set has less scatter. Both the mean error and a dispersion measure such as standard deviation are needed to describe performance. A microlitre is an exact unit relationship; an individual micropipette delivery is an experimental measurement.
Gilson 2026-11-05
Gilson is a laboratory-instrument manufacturer and the corporate author of Gilson Guide to Pipetting. In this bibliography, the name identifies the company, rather than an individual researcher. The guide documents small-volume liquid handling, especially the relation between a pipette setting, the liquid actually delivered, and experimental reproducibility.[ref]
In an air-displacement pipette, a piston changes the volume of an air cushion between the piston and the liquid. In a positive-displacement pipette, the piston acts directly on the liquid in the disposable assembly. The distinction helps explain why liquid properties and operating technique can affect delivery: the dial setting alone does not establish the delivered volume.[ref]
For example, repeated nominal water deliveries can be weighed and converted to volume using density and the calibration corrections. Their mean offset from measures a systematic error; variation between deliveries measures repeatability. The two answer different experimental questions: a tightly clustered set of deliveries can still have a biased mean.[ref]
Microlitre 2026-11-05
A microlitre, symbol , is one millionth of a litre, a unit of volume. Since a litre is exactly one cubic decimetre,
The prefix micro fixes the factor ; it does not describe how accurately an instrument can deliver that volume.[ref]
A sample occupies , the same volume as twenty cubes each one millimetre on a side. Delivering instead produces a 5% shortfall: a small absolute volume error can therefore matter greatly in a small-scale chemical reaction.
The microlitre is a volume unit, not a mass unit. To infer delivered volume from a balance measurement, use with the appropriate liquid density; careful calibration also accounts for effects such as air buoyancy and evaporation. Thus “one microlitre equals one milligram” is an approximation for water under particular conditions, not a unit identity. Gilson's calibration protocol illustrates repeated gravimetric measurements that separate systematic delivery error from scatter between deliveries.[ref]
Micropipette 2026-11-05
A laboratory micropipette is a pipette for transferring small measured quantities of liquid, commonly expressed in microlitres. In piston-operated models, the displacement mechanism matters. An air-displacement pipette leaves an air cushion between piston and sample, whereas a positive-displacement pipette places the piston directly against the liquid.[ref]
Weighing a delivered sample provides a concrete check. In a simplified calculation that neglects air buoyancy and evaporation, liquid volume follows from mass and density :
For an assumed , a delivered mass of corresponds to . Relative to a nominal setting, the deviation is . These are illustrative values, not calibration data.
A repeated series can distinguish a displaced mean from wide scatter: measurement accuracy and standard deviation describe different aspects of performance. Actual gravimetric checks account for temperature-dependent water density, air buoyancy and evaporation losses. The manufacturer's method uses a conversion factor and an evaporation correction, showing why a displayed volume alone does not establish the quantity delivered.[ref]
A positive-displacement pipette transfers liquid with a piston that contacts the sample directly, avoiding the intermediate air cushion of an air-displacement pipette. In the handheld designs described by the manufacturer, the capillary and piston form a disposable assembly. The mechanism can reduce effects associated with the expansion or contraction of trapped air.[ref]
This distinction is useful when handling liquids whose viscosity, density or volatility differs substantially from water. The instrument still measures a volume: a selected does not imply a fixed mass for every liquid. For an illustrative density of , the corresponding mass is by .
A gravimetric check therefore needs the appropriate density conversion. Direct displacement also does not make every delivery perfect: instrument condition, the fitted assembly and the test procedure remain relevant. Repeated measurements distinguish a systematic displacement of the mean from scatter in individual deliveries.[ref]