Intermediate measurement precision is measurement precision under declared intermediate precision conditions of measurement. The measurement procedure and location remain the same, measurements extend over a longer period, and specified changes can include operators, measuring systems or measurement calibration. It lies between a tightly restricted measurement repeatability study and a comparison across locations for measurement reproducibility.[ref]
International vocabulary of metrology by JCGM. VIM3 entries 2.22–2.23, definitions and entry 2.22 notes 1–2
For a concrete study, one laboratory could measure the same stable reference block each working day with a differential length comparator, taking four complete placements per day. Keep the procedure fixed, record temperature and operator, and retain the day labels. The labels let the study distinguish variation between days from rapid reading-to-reading variation.
An original measurement model makes that distinction explicit:Here is a shared daily offset, while varies between placements. Assume zero-mean independent random variables, independent also between days. A single measurement then has standard deviation . For stipulated and , this is , even though within-day scatter reflects only .
For four readings from one day, the daily arithmetic mean hasAveraging four such means from independent days reduces that modelled standard deviation to . Four same-day readings do not average away their shared daily offset. The numbers illustrate the model, not measured comparator performance. A real study must estimate its components from enough days and investigate trends: deterministic instrumental drift is not automatically an independent daily offset.
Articles by others on the same topic
There are currently no matching articles.