An intermediate precision condition of measurement is part of the specified conditions used to assess intermediate measurement precision. The measurement procedure and location remain common, while replicate measurements extend over a longer period. Declared changes can include operators, measuring systems, reference devices used for measurement calibration, or new calibrations.[ref] The record should state what changed as well as what was held fixed.
For an original length study, use the same laboratory, reference block and written comparator procedure in two sessions three weeks apart. Two operators each perform the complete remove-and-replace measurement cycle in both sessions. Record instrument identity, measurement calibration, temperature and settling conditions rather than treating “same laboratory” as a complete description.
Consider the following stipulated session means in millimetres:The operator difference is within either session, and the session difference is for either operator. In an additive model, those two comparisons distinguish the respective contributions because both operators appear in both sessions.
If operator A had measured only in session 1 and operator B only in session 2, the observed difference would combine the two changes. It could not be uniquely attributed to operator or elapsed time from that design alone. Repeated complete trials within each operator-session combination are still needed to assess random scatter; four chosen means do not establish a general precision value.
A measurement repeatability study keeps more conditions fixed over a short interval. Changing locations and measuring systems leads toward reproducibility conditions of measurement. The intermediate specification identifies the particular longer-term changes included in the precision study, rather than promising that every source of variation has been sampled.