Internal-standard-normalized process efficiency compares the overall preparation-and-response efficiency of an analyte with that of an internal standard. With matched corresponding amounts, let and be their mass spectrometry process efficiencies. The normalized quantity iswhere denotes extraction recovery and MF a mass spectrometry matrix factor. It combines recovery matching with the internal-standard-normalized matrix factor.
The underlying separation between recovery and matrix response is important in quantitative mass spectrometry.[ref]
Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS by B. K. Matuszewski and others. Abstract, experimental separation of matrix effects and recovery.
A ratio near one can conceal substantial losses affecting both signals.For original stipulated values, let analyte and standard each have recovery and matrix factor . Their process efficiencies are both , so the normalized value is 1 even though each observed response is only 30% of its matched neat reference. Now change only standard recovery to . Its process efficiency becomes and the normalized value becomes , despite identical matrix factors.
Addition timing changes the comparison. If the standard is added after the extraction stage, its recovery through that stage is effectively 1 by construction. Equal matrix responses can then cancel while the analyte's preparation loss remains. A standard added earlier can track more of the process, but equivalent behavior still requires evidence.
The desired property is a stable, appropriately calibrated analyte-to-standard response relationship. A constant nonunity normalized efficiency can be incorporated into a valid calibration; an unmodeled change across samples or levels can bias results. Report the reference, addition stage, quantities compared and blank correction rather than treating normalization as a guarantee of full material recovery.
Articles by others on the same topic
There are currently no matching articles.