A measuring chain is one successive signal path through a measuring system, beginning at a sensor and ending at an output element. The VIM illustrates this with both an electro-acoustic chain and a mechanical chain; a measuring chain need not be electronic.[ref]
Consider an original force measurement: a load cell produces an electric potential difference, an amplifier scales that signal, an analog-to-digital converter encodes it, and software converts the code to a force value. This is one measuring chain within the larger measuring system, which also includes its excitation supply, references and mounting arrangements. The load cell acts as a measuring transducer.
In an ideal linear model, let sensor response be , amplifier gain , and converter step per code increment. If the additive output offset is increments, the numerical output isThus implies . These are stipulated teaching values, not measured equipment performance.
The example exposes why measurement calibration must describe the relevant path. Changing amplifier gain changes the relation between output and force even if the load cell is untouched. A saturated stage can also discard information that later conversion cannot restore. Gain uncertainty, offsets, instrumental drift and finite conversion steps contribute at different points; treating the final display's increment as the whole measurement uncertainty overlooks those contributions.