John D. Ramboz 2026-11-05
John D. Ramboz is the first author, with Oskars Petersons, of NIST's 1991 account of a laboratory current transformer calibration service. The report describes ratio and phase measurements, their uncertainties and quality-control procedures for electrical measurement work.[ref] It is a primary technical record of the service and its measurement conventions.
The separation of magnitude and phase has a direct consequence for electric power measurements. Consider an original sinusoidal model with root-mean-square electric potential difference , secondary current magnitude , true current ratio and primary-current lag . True active power is
If the secondary leads the primary by , simply multiplying the secondary signal by nominal ratio gives
provided . This uses the lead-positive current-transformer phase displacement convention stated in the report.[ref]
With exact magnitude conversion and small in radians, the relative power error is approximately . For stipulated rad and , that term is about . At , its first-order value vanishes. Thus the same phase error can have different effects at different power factors.
This derivation is an explanatory example, not a reported performance test or an attribution of the calculation to Ramboz. It explains why a calibration report needs operating conditions, a sign convention and measurement uncertainty, in addition to a nominal transformer ratio. Precision in the current magnitude does not remove phase-dependent power error.
Oskars Petersons 2026-11-05
Oskars Petersons worked on electrical metrology at the National Institute of Standards and Technology. With John D. Ramboz, Petersons coauthored the 1991 report describing NIST's laboratory current transformer calibration service. The report treats current ratio and current-transformer phase displacement as distinct measurement results.[ref]
A current transformer allows a secondary electric current to represent a larger or smaller primary current. Its marked ratio is a nominal conversion, whereas the measured ratio can differ. The report expresses the measured ratio as the marked ratio multiplied by the current-transformer ratio correction factor, and uses a secondary-leading sign convention for the reported phase displacement.[ref]
An original calculation illustrates the distinction. A marked ratio of A: A gives . If the current-transformer ratio correction factor is , the measured magnitude ratio is . A secondary reading of 5 A then represents 200.04 A under those stated conditions, rather than exactly 200 A.
Magnitude alone is not a full alternating-current description. If the secondary phasor leads the primary by and their magnitude ratio is , a consistent convention is
The magnitude correction and phase rotation address different errors. A measurement system can reproduce one accurately while requiring correction of the other. The joint report's calibration circuits, including compensated current comparators, give these distinctions an apparatus-level setting rather than leaving them as ideal-transformer identities.[ref]