Minimum-noise weights under baseline moment constraints choose a linear pulse estimator with the smallest variance among those that preserve a specified signal and remove specified backgrounds. Write the reading vector as , where is a known signal template, the columns of contain baseline shapes, and the error covariance matrix is positive definite. For , requireColumns in implement the moment conditions for baseline-drift cancellation. The noise cost of polynomial baseline cancellation depends on which feasible weights are chosen.
If the constraint columns are independent, Lagrange multipliers giveFor any feasible perturbation , and , so the variance increases by . This proves the minimum and connects the estimator to generalized least squares. The covariance propagation is the full quadratic form, not a sum that discards correlations.[ref]
For an original one-bin pulse example, use five equally spaced times , a signal present only at the centre, and independent errors of variance . To remove just constant and linear backgrounds, the optimal dimensionless weights are . Their variance is , lower than for the centre and its two nearest references alone. Multiplying the amplitude estimate by a bin duration gives energy variance . Extra uncontaminated references can reduce noise when the cancelled background family stays the same.
The inverse requires genuinely independent constraints. Constraint redundancy in baseline-noise minimization calls for removing redundant columns and checking feasibility; a signal template inside the nuisance span cannot be preserved while that span is rejected. NIST's before/after pulse baselines provide the physical measurement context.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. §3.5, printed p. 8; §7.2, Leakage Effect
The five-point design is an original idealization and assumes no pulse signal in its reference samples.The noise cost of polynomial baseline cancellation is the change in random measurement uncertainty when extra background trends are removed. For a linear pulse estimate and reading-error covariance matrix ,The moment conditions for baseline-drift cancellation constrain to reject specified polynomial backgrounds. They do not make the remaining random error vanish.[ref]
For an original comparison, assume a pulse occupies only the central sample and the other samples contain background alone. Three samples at with weights cancel constant and linear backgrounds. Five samples at can instead useThese weights preserve the central signal and have zero moments of orders . For independent reading errors of equal variance , the three-sample variance is , while the five-sample variance isWith and , the standard uncertainties are approximately and . The second design removes quadratic and cubic drift as well, but its larger reference weights increase noise. This compares two specified estimators; it is not a claim that collecting more background data must worsen precision.
NIST's pulse-energy apparatus used before/after baseline measurements for background and thermal-drift correction.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. §3.5, printed p. 8; §7.2, Leakage Effect
The polynomial comparison above is an original model for choosing such measurements. Choosing minimum-noise weights under baseline moment constraints can exploit additional reference samples and their measured correlations. Pulse leakage into a reference or drift outside the cancelled polynomial family still creates bias.