Define the retard ratio using the kinetic energy at the analyser entrance before the deliberate retardation: . Initially . With the original excitation, a carbon photoelectron arrives with . ThusThe new deliberate retardation is . For fixed slits, angular acceptance and hemisphere radii, the geometrical fractional energy width is approximately constant, so the absolute analyser width grows in proportion to pass energy: about 2.03 times poorer absolute energy resolution. Counts may improve, but closely separated chemical shifts become harder to resolve. An alternative definition gives 0.114 rather than 1.114 for the same physical setting and leads to the same final pass energy when used consistently.