Randall M. Schoonover Created 2026-10-14 Updated 2026-11-05
Randall M. Schoonover investigated precision measurements of mass and mass density. A coauthored study used an electronic force balance to weigh a silicon crystal in air and water, applying buoyancy and measurement calibration corrections. The reported crystal density agreed with its accepted value within 2.4 parts per million.[ref]
The method obtains mass density from the two buoyant-force readings and known fluid densities; it does not require an independent measurement of the crystal's volume.[ref]
NISTIR 5375, issued by National Institute of Standards and Technology in February 1994, reprinting ISA paper 93-091, printed pp. 299–320. The reprinted paper bears an ISA 1993 copyright; this is distinct from the NIST report publication.[ref]
Cited passages: printed pp. 313–316, Pycnometer Calibration, Figure 5 and Pycnometer Application: calibration and correction for bath-versus-weighing temperature; printed p. 315: tin-oxide charge precaution, no detected charge, and pycnometer data not yet analyzed.